Skip to main content
. 2012 Apr 24;7:63. doi: 10.1186/1748-717X-7-63

Table 1.

Patient positioning errors prior to cone-beam CT based image-guidance (IG) and immediately following image-guided radiosurgery (IG-RS)

  Prior IG (n = 98) Post IG-RS (n = 64)
 
Average
StDev
Max
Average
StDev
Max
LR [mm]
0.1
2.1
10.3
0.1
0.6
1.8
SI [mm]
−0.8
1.7
5.4
−0.3
0.8
3.0
AP [mm]
−2.7
2.0
9.9
−0.2
0.4
1.4
3D vector [mm]
3.9
1.9
11.9
0.9
0.6
3.0
Max Rotation [°] 1.7 0.8 4.0 0.6 0.5 3.0