Dose distributions to the target with simulation of 0 mm, 1 mm and 2 mm safety margins 1) in the treatment plan (plan), 2) in the scenario of radiosurgery without image-guidance (No IG), 3) in the scenario of radiosurgery after image-guided correction of translational errors only and not rotations (IG trans) and 4) simulating errors observed immediately following image-guided radiosurgery (Post IG-RS)

Safety margin   Plan No IG IG trans Post IG-SRS
0 mm
Absolute CI
0.96
±0.06
0.72
±0.19
0.96
±0.06
0.94
±0.07
 
<95% planned CI
 
91%
3%
14%
1 mm
Absolute CI
1.00
±0.01
0.82
±0.19
1.00
±0.01
0.99
±0.01
 
<95% planned CI
 
70%
0%
0%
2 mm
Absolute CI
1.00
±0.04
0.90
±0.17
1.00
±0.04
1.00
±0
  <95% planned CI   40% 0% 0%

Absolute values of the coverage index (average ± standard deviation) and percent of the patients with <95% planned target coverage are summarized.