Published in final edited form as: J Pharm Biomed Anal. 2012 May 23;70:320–329. doi: 10.1016/j.jpba.2012.05.008

Table 4.

Intra-day and inter-day precision and accuracy of the method for umbilical plasma and placental tissue.

Bio-matrix Analytea Addedb
concentration
Intra-day (n=6) Inter-day (n=3)

Mean
concentration
obtainedb
Mean
accuracyc
(%)
Precision
(RSD, %)
Mean
concentration
obtainedb
Mean
accuracyc
(%)
Precision
(RSD, %)
Umbilical cord plasma BUP 0.250 0.220 88.1 11.3 0.237 94.7 10.4
0.625 0.553 88.5 5.3 0.577 92.4 4.4
62.5 59.2 94.7 6.2 60.8 97.3 4.4
188 190 101.5 4.1 186.6 99.5 1.8
OH-BUP 0.715 0.702 98.2 5.7 0.680 95.0 7.4
1.79 1.77 98.9 5.7 1.774 99.2 1.8
179 166 92.9 5.5 172.8 96.7 3.6
536 512 95.5 2.2 519.0 96.8 2.0
EB 0.245 0.239 97.7 5.3 0.234 95.7 6.4
0.612 0.571 93.2 3.8 0.587 95.8 4.3
61.2 57.1 93.3 6.0 58.8 95.9 3.1
184 182 98.9 4.9 180.2 98.1 1.7
TB 0.365 0.347 95.0 8.3 0.344 94.3 2.3
0.912 0.860 94.2 4.8 0.873 95.7 2.1
91.2 85.9 94.2 7.4 89.1 97.7 4.1
274 266 97.1 5.0 266.4 97.3 1.1
Placental tissue BUP 0.312 0.320 102.4 9.6 0.301 96.4 9.3
0.625 0.605 96.8 8.8 0.605 96.8 7.7
62.5 64.7 103.6 4.9 64.2 102.7 5.6
188 185 98.5 8.6 185 98.8 7.4
OH-BUP 0.895 0.811 90.7 12.0 0.839 93.8 12.5
1.79 1.78 99.4 5.1 1.76 98.3 6.1
179 188 105.1 4.1 182 101.8 6.4
537 522 97.3 8.8 522 97.2 7.2
EB 0.305 0.273 89.6 15.2 0.279 91.6 13.2
0.610 0.597 97.9 8.8 0.587 96.3 8.3
61.0 63.2 103.6 3.8 61.8 101.3 5.7
183 181 98.7 8.4 180 98.5 7.4
TB 0.918 0.856 93.3 11.5 0.858 93.5 11.2
1.84 1.85 100.6 6.5 1.82 99.1 6.5
184 193 105.3 4.1 190 103.5 5.4
550 524 95.3 7.7 526 95.5 9.6
a

BUP=buproipion, OH-BUP=hydroxybupropion, EB=erythrohydrobupropion, TB=threohydrobupropion.

b

The unit of analyte in umbilical cord plasma is ng/mL; the unit of analyte in placental tissue is ng/g.

c

accuracy=(obtained concentration/added concentration)×100