Skip to main content
. Author manuscript; available in PMC: 2012 Nov 7.
Published in final edited form as: Phys Med Biol. 2012 Sep 12;57(19):6025–6046. doi: 10.1088/0031-9155/57/19/6025

Figure 11.

Figure 11

Results of measured (symbols) and calculated (lines) light fluence rate at each detector locations (D1 – D6) according to Fig. 3 during PDT treatment. Each plot is the sum of the light fluence accumulated from the four adjacent channels. Each of the 12 CDFs has a uniform source strength of 100 mW/cm.