Published in final edited form as: Phys Med Biol. 2012 Sep 12;57(19):6025–6046. doi: 10.1088/0031-9155/57/19/6025

Table 3.

Results for heterogeneous phantom. The first three columns list the known values and position of μa for the phantom background and three heterogeneities (numbered 1, 2, 3, clockwise starting from upper right corner) according to Fig. 3. Column four lists the mean and standard deviation of the measured μa for the background and the three heterogeneities. The last column specifies the percentage difference between the measured and the true μa value.

Description Region specification True μa values Uniform Fit In-air scan fit

Measured μa values % diff Max % diff Measured μa values % diff Max % diff

Background D2, S3 (Fig. 8d) 0.2 0.212±0.026 6% 26% 0.212±0.026 6% 26%
Heterogeneity 1 D3, S4 (Fig. 8f) 0.5 0.381±0.024 24% 33% 0.390±0.022 22% 30%
Heterogeneity 2 D5, S11 (Fig. 8h) 0.1 0.130±0.030 30% 69% 0.127±0.029 27% 57%
Heterogeneity 3 D1, S1 (Fig. 8b) 0.9 0.654±0.085 27% 37% 0.731±0.146 19% 37%