TABLE 5.

Results of clinical evaluation of diagnostic Ortho/ECi assay and Abbott EIA following resolution of discrepant samplesa

Sample type Ortho/ECi result
Abbott/EIA result
Positive Negative Positive Negative
Anti-HCVb
    Positive 175 0 175 2
    Negative 6 311 0 315
HBsAgb
    Positive 149 4 152 5
    Negative 0 316 1 311
Anti-HBsAg
    Positive 241 3 236 3
    Negative 17 219 8 233
a

Concordant positive or negative samples between the two assays and discrepant samples that were confirmed as true positive or negative based on supplemental tests and chart review were employed to determine false-positive or -negative Ortho/ECi and Abbott EIA results.

b