. 2012 Aug 3;12:174. doi: 10.1186/1471-2334-12-174

Table 1.

Ability of the Nanoclave Cabinet to disinfect non-critical patient care items

  minimum and maximum log10reduction after exposure to two 30-second UV-C cyclesa
Blood pressure gauge (n = 8) Patient call button (n = 8) Infusion pump (n = 5) Tympanic thermometer (n = 9) Oximeter (base unit) (n = 7) Computer keyboard/mouse (n = 4) TV remote control (n = 4) Blood pressure cuff (n = 6)
Pathogen b
min
max
min
max
min
max
min
max
min
max
min
max
min
max
min
max
MRSA
4.40
>5.29
>4.74
>5.17
>4.94
>5.08
2.16
>5.45
>5.25
>5.48
>4.97
>5.10
>5.05
>5.32
1.93
>5.13
VRE
>5.11
>5.23
>5.05
>5.21
>4.93
>5.30
1.49
>5.44
>4.91
5.27
4.28
>5.03
4.93
>5.16
2.13
>5.00
A. baumannii
3.44
>5.54
5.32
>5.59
>5.39
>5.56
2.29
>5.64
>5.13
>5.48
4.90
>5.74
>5.33
>5.75
3.46
>5.39
Kleb. pneum 2.76 >5.19 >4.84 >5.25 4.04 5.07 1.02 >5.11 >4.33 >5.08 >5.11 >5.21 >5.05 >5.12 3.24 >5.34

ainitial inoculum: 106 cfu/cm2.

btest organism suspended in ¼-strength Ringer’s solution.