Skip to main content
. 2012 Aug 3;12:174. doi: 10.1186/1471-2334-12-174

Table 2.

Comparative performance of the Nanoclave Cabinet and antimicrobial wipes when used to disinfect patient care items

  Mean (± SD) log10reductiona(n = 3)
Nanoclave
Antimicrobial wipes
(2 × 30 sec) ‘poor’ wiping ‘moderate' wiping ‘thorough’ wiping
Tympanic thermometer
display panel
>5.49
4.45 ± 1.03
5.04 ± 0.60
>5.39
infra-red sensor
2.29 ± 0.93
1.94 ± 0.03
1.96 ± 0.25
2.14 ± 0.14
plastic lid
>5.64
>5.40
>5.40
>5.40
probe receptor
4.55 ± 1.07
5.16 ± 0.17
>5.48
>5.48
earpiece holder
3.44 ± 0.13
5.25 ± 0.40
5.28 ± 0.35
>5.49
Blood pressure cuff
Velcro (hook)
3.60 ± 0.98
1.91 ± 0.07
2.42 ± 0.14
2.66 ± 0.09
Velcro (loop)
4.28 ± 0.96
1.50 ± 0.03
2.26 ± 0.19
2.67 ± 0.06
inner cuff surface
3.46 ± 1.47
1.73 ± 0.09
2.65 ± 0.08
2.30 ± 0.08
pump
>5.39
2.38 ± 0.15
2.90 ± 0.08
>5.60
pump tubing >5.07 2.75 ± 0.40 3.94 ± 1.22 >5.33

a initial inoculum: 106 cfu/cm2.