Skip to main content
. 2006 Dec 13;24(1):5–9. doi: 10.1007/s10815-006-9088-3

Table 1.

Comparison of micro-environment recovery

Front-load
Top-load conventional
mini-incubator incubator
Temperature 4.9 ± 0.5 min* 31.5 ± 2.9 min
O2 3.0 ± 0 min* 7.8 ± 0.9 min

Note. Time-lapse for recovery after opening the door for 5 s.

*P<0.01 (mini-incubator vs. conventional incubator).