Table 1.

Measurements of prostate blood flow (PBF) and estimated measurement errors due to random and temporal noises

Subject No. Age
(years)
Volume
(cm3)
PBF
(mL/100 cm3/min)
Random Error
(mL/100 cm3/min)
Random Error
/PBF (%)
Temporal Error
(mL/100 cm3/min)
Temporal Error
/PBF (%)
1 29 12.3 33.9 0.68 2.1 4.8 14.9
2 29 11.2 22.5 0.91 4.3 4.5 21.1
3 39 10.0 31.2 1.59 5.4 2.5 8.4
4 57 17.8 31.7 0.81 2.7 7.5 24.9
5 65 16.8 16.4 0.49 3.1 5.7 36.6
Mean 44 13.6 25.8 0.90 3.5 5.0 21.2
Standard
Deviation
16 3.5 7.1 0.42 1.3 1.8 10.7