Table 5.

Calculated positive and negative predictive values for FilmArray RP and DFCI/BWH clinically indicated testing

Sample typea Test Calculated positive predictive value (CI)b Calculated negative predictive value (CI)b
Allc DFCI/BWH 1.00 (0.79, 1.00) 0.84 (0.74, 0.91)
FilmArray RP 0.86 (0.68, 0.96) 1.00 (0.95, 1.00)
BALc DFCI/BWH 1.00 (0.37, 1.00) 0.90 (0.78, 0.97)
FilmArray RP 0.89 (0.52, 1.00) 1.00 (0.93, 1.00)
NPA DFCI/BWH 1.00 (0.74, 1.00) 0.71 (0.49, 0.87)
FilmArray RP 0.85 (0.62, 0.97) 1.00 (0.81, 1.00)
a

BAL, bronchoalveolar lavage; NPA, nasopharyngeal aspirate.

b

The estimated positive and negative predictive values did not change much when the second sample obtained from the three patients who underwent testing twice was excluded. CI, confidence interval.

c

These estimates excluded the three samples in which parainfluenza virus 2 and parainfluenza virus 3 were detected by clinical testing but not by FilmArray RP assay, since verification testing did not test for either of these pathogens.