Table 1. Intrasession Repeatability of 8 Different Devices in Measuring Flat Keratometry (N = 35).

Device Session Mean (D) ± SD Sw (D) 2.77 Sw (D) COV (%) ICC
Tomey RC 1st 42.85±1.26 0.04 0.10 0.08 0.999
2nd 42.85±1.27 0.03 0.09 0.08 0.999
Topcon KR 1st 42.90±1.30 0.05 0.15 0.13 0.998
2nd 42.91±1.27 0.06 0.17 0.14 0.998
IOLMaster 1st 42.97±1.27 0.03 0.09 0.07 0.999
2nd 42.99±1.27 0.04 0.10 0.08 0.999
EyeSys Vista 1st 42.65±1.21 0.13 0.36 0.30 0.989
2nd 42.64±1.19 0.10 0.29 0.24 0.992
Medmont 1st 43.05±1.29 0.07 0.20 0.17 0.997
2nd 43.06±1.27 0.07 0.20 0.17 0.997
Topolyzer 1st 42.90±1.28 0.10 0.27 0.23 0.994
2nd 42.91±1.25 0.10 0.27 0.23 0.994
Pentacam 1st 42.90±1.26 0.08 0.21 0.18 0.996
2nd 42.90±1.26 0.07 0.20 0.17 0.997
Sirius 1st 42.97±1.26 0.10 0.27 0.23 0.994
2nd 43.01±1.28 0.08 0.22 0.18 0.996

D = diopter, SD = standard deviation, Sw = within-subject standard deviation, COV = within-subject coefficient of variation, ICC = intraclass correlation coefficient.