Skip to main content
. 2012 Sep 25;7(9):e45607. doi: 10.1371/journal.pone.0045607

Table 3. Intrasession Repeatability of 8 Different Devices in Measuring Mean Keratometry (N = 35).

Device Session Mean (D) ± SD Sw (D) 2.77 Sw (D) COV (%) ICC
Tomey RC 1st 43.36±1.31 0.03 0.08 0.06 1.000
2nd 43.36±1.33 0.03 0.07 0.06 1.000
Topcon KR 1st 43.32±1.34 0.06 0.16 0.13 0.998
2nd 43.33±1.33 0.06 0.17 0.14 0.998
IOLMaster 1st 43.47±1.32 0.02 0.06 0.05 1.000
2nd 43.48±1.32 0.03 0.07 0.06 1.000
EyeSysVista 1st 43.09±1.26 0.11 0.32 0.26 0.992
2nd 43.09±1.26 0.10 0.29 0.24 0.993
Medmont 1st 43.57±1.33 0.07 0.18 0.15 0.999
2nd 43.60±1.33 0.06 0.17 0.14 0.998
Topolyzer 1st 43.37±1.34 0.11 0.29 0.24 0.994
2nd 43.37±1.32 0.10 0.28 0.23 0.994
Pentacam 1st 43.38±1.31 0.07 0.18 0.15 0.998
2nd 43.38±1.32 0.06 0.17 0.14 0.998
Sirius 1st 43.44±1.33 0.09 0.25 0.21 0.996
2nd 43.48±1.34 0.08 0.21 0.18 0.997

D = diopter, SD = standard deviation, Sw = within-subject standard deviation, COV = within-subject coefficient of variation, ICC = intraclass correlation coefficient.