Table 4. Comparison of the Flat Keratometry between 8 Different Devices.
Devices | Mean Difference(D) ± SD | P Value | 95% LoA (D) |
Tomey-Topcon | −0.05±0.09 | <.01 | −0.226 to 0.134 |
Tomey-IOLMaster | −0.12±0.08 | <.01 | −0.278 to 0.044 |
Tomey-EyeSys | 0.20±0.11 | <.01 | −0.010 to 0.41 |
Tomey-Medmont | −0.20±0.15 | <.01 | −0.480 to 0.090 |
Tomey-Topolyzer | −0.05±0.09 | <.01 | −0.231 to 0.133 |
Tomey-Pentacam | −0.05±0.09 | <.01 | −0.236 to 0.132 |
Tomey-Sirius | −0.12±0.11 | <.01 | −0.330 to 0.09 |
Topcon-IOLMaster | −0.07±0.09 | <.01 | −0.246 to 0.103 |
Topcon-EyeSys | 0.24±0.14 | <.01 | −0.02 to 0.51 |
Topcon-Medmont | −0.15±0.14 | <.01 | −0.42 to 0.12 |
Topcon-Topolyzer | 0.00±0.10 | .851 | −0.196 to 0.19 |
Topcon-Pentacam | −0.01±0.11 | .735 | −0.22 to0.21 |
Topcon-Sirius | −0.08±0.12 | <.01 | −0.31 to0.15 |
IOLMaster-EyeSys | 0.32±0.12 | <.01 | 0.07 to 0.56 |
IOLMaster-Medmont | −0.08±0.15 | <.01 | −0.37 to 0.21 |
IOLMaster-Topolyzer | 0.07±0.08 | <.01 | −0.082 to 0.219 |
IOLMaster-Pentacam | 0.07±0.11 | <.01 | −0.16 to 0.29 |
IOLMaster-Sirius | 0.00±0.11 | .859 | −0.22 to 0.22 |
EyeSys-Medmont | −0.40±0.17 | <.01 | −0.73 to −0.06 |
EyeSys-Topolyzer | −0.25±0.13 | <.01 | −0.5 to 0 |
EyeSys-Pentacam | −0.25±0.11 | <.01 | −0.47 to 0.03 |
EyeSys-Sirius | −0.32±0.12 | <.01 | −0.56 to −0.08 |
Medmont-Topolyzer | 0.15±0.15 | <.01 | −0.14 to 0.44 |
Medmont-Pentacam | 0.14±0.14 | <.01 | −0.14 to 0.43 |
Medmont-Sirius | 0.08±0.17 | .014 | −0.26 to 0.42 |
Topolyzer-Pentacam | 0.00±0.11 | .872 | −0.23 to 0.22 |
Topolyzer-Sirius | −0.07±0.12 | <.01 | −0.31 to0.17 |
Pentacam-Sirius | −0.07±0.10 | <.01 | −0.263 to 0.126 |
D = diopter, SD = standard deviation, LoA = limits of agreement.