. 2012 Sep 25;7(9):e45607. doi: 10.1371/journal.pone.0045607

Table 4. Comparison of the Flat Keratometry between 8 Different Devices.

Devices Mean Difference(D) ± SD P Value 95% LoA (D)
Tomey-Topcon −0.05±0.09 <.01 −0.226 to 0.134
Tomey-IOLMaster −0.12±0.08 <.01 −0.278 to 0.044
Tomey-EyeSys 0.20±0.11 <.01 −0.010 to 0.41
Tomey-Medmont −0.20±0.15 <.01 −0.480 to 0.090
Tomey-Topolyzer −0.05±0.09 <.01 −0.231 to 0.133
Tomey-Pentacam −0.05±0.09 <.01 −0.236 to 0.132
Tomey-Sirius −0.12±0.11 <.01 −0.330 to 0.09
Topcon-IOLMaster −0.07±0.09 <.01 −0.246 to 0.103
Topcon-EyeSys 0.24±0.14 <.01 −0.02 to 0.51
Topcon-Medmont −0.15±0.14 <.01 −0.42 to 0.12
Topcon-Topolyzer 0.00±0.10 .851 −0.196 to 0.19
Topcon-Pentacam −0.01±0.11 .735 −0.22 to0.21
Topcon-Sirius −0.08±0.12 <.01 −0.31 to0.15
IOLMaster-EyeSys 0.32±0.12 <.01 0.07 to 0.56
IOLMaster-Medmont −0.08±0.15 <.01 −0.37 to 0.21
IOLMaster-Topolyzer 0.07±0.08 <.01 −0.082 to 0.219
IOLMaster-Pentacam 0.07±0.11 <.01 −0.16 to 0.29
IOLMaster-Sirius 0.00±0.11 .859 −0.22 to 0.22
EyeSys-Medmont −0.40±0.17 <.01 −0.73 to −0.06
EyeSys-Topolyzer −0.25±0.13 <.01 −0.5 to 0
EyeSys-Pentacam −0.25±0.11 <.01 −0.47 to 0.03
EyeSys-Sirius −0.32±0.12 <.01 −0.56 to −0.08
Medmont-Topolyzer 0.15±0.15 <.01 −0.14 to 0.44
Medmont-Pentacam 0.14±0.14 <.01 −0.14 to 0.43
Medmont-Sirius 0.08±0.17 .014 −0.26 to 0.42
Topolyzer-Pentacam 0.00±0.11 .872 −0.23 to 0.22
Topolyzer-Sirius −0.07±0.12 <.01 −0.31 to0.17
Pentacam-Sirius −0.07±0.10 <.01 −0.263 to 0.126

D = diopter, SD = standard deviation, LoA = limits of agreement.