Skip to main content
. 2012 Nov;70(6):523–528. doi: 10.1016/j.jpba.2012.05.028

Table 2.

Intra- and inter-day precision and accuracy data for assays of RIF in human plasma.

Inter-day (n = 18)
Intra-day (n = 6)
Mean conc.a (ng/mL) CV % Accuracy % Mean conc. (ng/mL) CV % Accuracy %
LLOQ (25 ng/mL) 23.23 ± 1.34 5.78 92.93 24.64 ± 1.51 6.16 98.58
LQC (75 ng/mL) 79.07 ± 6.13 7.75 105.43 82.59 ± 1.52 1.85 110.12
MQC (600 ng/mL) 641.33 ± 37.74 5.88 106.89 649.10 ± 16.81 2.59 108.18
HQC (4800 ng/mL) 5156.19 ± 268.85 5.21 107.42 4846.14 ± 109.91 2.27 100.96
HLOQ (6400 ng/mL) 6610.76 ± 223.44 3.38 103.29 6590 ± 372.57 5.65 102.98
a