Skip to main content
. 2012 Oct;14(10):877–882. doi: 10.1089/dia.2012.0079

Table 1.

Percentages of Glucose Measurements in Different Ranges Comparing Continuous Glucose Monitoring and Reference Glucose Measurements

 
Overall
Daytime
Nighttime
  CGM Reference CGM Reference CGM Reference
All values (n) 52,146 5,195 27,944 2,830 24,202 2,365
 71–180 mg/dL (%) 59 60 58 56 61 65
 >180 mg/dL (%) 28 31 35 36 21 24
 >250 mg/dL (%) 9 10 10 12 8 8
 ≤70 mg/dL (%) 12 9 8 7 17 11
 ≤60 mg/dL (%) 6 4 4 3 9 5
 Mean glucose (mg/dL) 148 154 159 164 134 142
 Area under the curve 70 mg/dL 1.31 0.77 0.80 0.59 1.89 0.98
 Low Blood Glucose Index 2.4 1.8 1.6 1.4 3.4 2.2
 Area under the curve 180 mg/dL 16.04 19.16 19.18 23.83 12.41 13.58
 High Blood Glucose Index 6.6 7.5 7.8 8.9 5.1 5.8
 Glucose CV (%) 47 47 43 46 50 47
Limited to paired values (n) 5,206 5,206 2,950 2,950 2,256 2,256
 71–180 mg/dL (%) 60 60 58 56 63 65
 >180 mg/dL (%) 31 31 37 36 23 23
 >250 mg/dL (%) 9 10 10 12 8 7
 ≤70 mg/dL (%) 9 9 6 7 14 12
 ≤60 mg/dL (%) 5 4 3 3 7 5
 Mean glucose (mg/dL) 153 154 163 164 139 141
 Area under the curve 70 mg/Dl 0.97 0.83 0.60 0.61 1.45 1.11
 Low Blood Glucose Index 1.9 1.8 1.3 1.4 2.7 2.4
 Area under the curve 180 mg/dL 17.38 19.01 20.75 23.57 12.97 13.06
 High Blood Glucose Index 7.1 7.4 8.3 8.8 5.5 5.6
 Glucose CV (%) 45 47 42 46 47 47

CGM, continuous glucose monitoring; CV, coefficient of variation (SD/mean).