. 2012 Jun 6;470(11):3083–3093. doi: 10.1007/s11999-012-2400-x

Table 7.

Studies with minimum 5-year followup comparing wear, osteolysis, and clinical function score between XLPE and conventional PE

Study Design THA type Femoral head XLPE/PE type Mean followup (years)* Annual linear steady state wear (mm/year) Wear assessment method Osteolysis Function score
Capello et al. [5] (2011) Retrospective matched control Uncemented CoCr 28 mm Crossfire®/Stryker® UHMWPE 7.5 (6–10.2) XLPE: 0.031 (0.01)
PE: 0.14 (0.08)
Image analysis
Livermore et al. [32]
XLPE no osteolysis NR
Garcia-Rey et al. [17] (2008) Randomized Uncemented CoCr 28 mm Durasul®/Sulene® 5.5 (5–7.7) XLPE: 0.006 (0.007)
PE: 0.036 (0.01)
Image analysis
Dorr and Wan [13]
No osteolysis No difference
Fukoi et al. [16] (2011) Consecutive series Uncemented/hybrid Zirconia ceramic 26 mm Longevity®/Zimmer® UHMWPE 5.4 (5–6.4) XLPE: 0.0068 (NR)
PE: 0.01 (NR)
Image analysis
Easy Scale for Orthopaedics v 3.1.2. (Nexis, Tokyo, Japan)
XLPE no osteolysis NR
Leung et al. [31] (2007) Randomized Uncemented CoCr 28 mm Marathon®/Enduron® 5 XLPE: 0.01 (NR)
PE: 0.2 (NR)
Image analysis
Martell and Berdia [33]
XLPE reduced osteolysis NR
McCalden et al. [34] (2009) Randomized Hybrid CoCr 28 mm Longevity®/Zimmer® UHMWPE 6.8 (5.6–7.6) XLPE: 0.003 (0.05)
PE: 0.051 (0.04)
Image analysis
Martell and Berdia [33]
NR No difference
Olyslaegers et al. [39] (2008) Retrospective matched control Hybrid CoCr 28 mm Longevity®/Zimmer® UHMWPE 5 XLPE: 0.05 (NR)
PE: 0.1 (NR)
Image analysis
Martell and Berdia [33]
No difference No difference
Rayadhyaksha et al. [43] (2009) Retrospective matched control Uncemented/hybrid CoCr 28 mm Crossfire®/Stryker® UHMWPE 6.1 (5–8.1) XLPE: 0.054 (0.047)
PE: 0.13 (0.076)
Image analysis
Martell and Berdia [33]
NR No difference
Thomas et al. [52] (2011) Randomized Hybrid CoCr 28 mm Longevity®/Zimmer® UHMWPE NR (7–7.8) XLPE: 0.005 (0.02)
PE: 0.037 (0.03)
RSA
Selvik [50]
No difference No difference
Current study Randomized Cemented CoCr 28 mm Durasul®/Sulene® 10 XLPE: 0.005 (0.01)
PE: 0.055 (0.05)
RSA
Selvik [50]
No difference No difference

* Values are expressed as mean, with range in parentheses; values are expressed as mean, with SD in parentheses; SD was calculated by 0.5 x 95% CI x √N, where N = number of participants in group; XLPE = highly crosslinked polyethylene; PE = polyethylene; NR = not reported; RSA = radiostereometry.