Table 1.

Data acquisition parameters for non-slice selective 2D and 3D UTE imaging.

UTE Imaging Protocol FOV [cm2] Reconstructed Matrix In-plane Resolution [mm2] Slice Thickness [mm] Sampling Bandwidth [kHz] Flip Angle [°] TE [ms] TR [ms] Number of Projections Total Scan Time [min]
2D UTE T2* bi-component analysis 8×8 256×256 0.31×0.31 - 125 10 0.008, 0.1, 0.2, 0.3, 0.4, 0.6, 0.8, 1, 1.2, 1.4, 1.7, 2, 2.5, 3, 3.5, 4, 5, 6, 7, 8 200 411 27
3D UTE total bone water measurement 15×15 256×256×256 0.58×0.58 0.58 250 10 0.008 300 50000 250