Table 3.

Comparison between 3D QCA and OCT in assessing lumen size

OCT 3D QCA Difference (95% CI) Intra-observer variabilitya Inter-observer variabilitya
Positions, n = 541
 Short diameter (mm) 2.70 ± 0.65 2.57 ± 0.61 0.14(0.11–0.16) 0.000 ± 0.013 0.003 ± 0.029
 Long diameter (mm) 3.11 ± 0.72 2.80 ± 0.62 0.30 (0.27–0.33) 0.003 ± 0.024 0.006 ± 0.035
 Lumen area (mm2) 7.01 ± 3.28 5.93 ± 2.66 1.07 (0.95–1.20) 0.002 ± 0.039 0.021 ± 0.059
Vessels, n = 40
 Short diameter (mm) 2.71 ± 0.46 2.57 ± 0.43 0.14 (0.09–0.19)
 Long diameter (mm) 3.11 ± 0.52 2.81 ± 0.45 0.30 (0.24–0.37)
 Lumen area (mm2) 7.02 ± 2.34 5.94 ± 1.91 1.08 (0.80–1.37)

aObserver variability was calculated from 165 positions from the first 10 vessels. CI Confidence interval; †  P < 0.001