TABLE 4.

Agreement and Correlation between RTVue and Stratus Measurements

Healthy Participants Glaucoma Patients


Optic Disc Parameters Mean Difference (range) r2 (P) Mean Difference (range) r2 (P)
TEMP RNFL (µm) 7.5 (−13.5 to 28.5) 0.44 (<.001) 5.4 (−7.5 to 18.2) 0.75 (<.001)
SUP RNFL (µm) −3.6 (−23.8 to 16.7) 0.6 (<.001) 2.1 (−20.1 to 24.2) 0.71 (<.001)
NAS RNFL (µm) −3.6 (−25.9 to 18.8) 0.58 (<.001) 0.3 (−18.9 to 19.5) 0.65 (<.001)
INF RNFL (µm) 6.4 (−13.8 to 26.6) 0.74 (<.001) 10.8 (−7 to 28.5) 0.83 (<.001)
AVG RNFL (µm) 1.7 (−8.6 to 12) 0.81 (<.001) 4.6 (−5.6 to 14.8) 0.86 (<.001)
Disc area (mm2) −0.4 (−1 to 0.2) 0.31 (<.001) −0.4 (−0.8 to 0) 0.82 (<.001)
Cup area (mm2) −0.1 (−0.7 to 0.6) 0.75 (<.001) −0.1 (−0.6 to 0.4) 0.84 (<.001)
Rim area (mm2) −0.3 (−1.1 to 0.4) 0.22 (.002) −0.3 (−0.7 to 0.1) 0.67 (<.001)
Rim volume (mm3) −0.3 (−0.8 to 0.1) 0.78 (<.001) −0.2 (−0.5 to 0.2) 0.76 (<.001)
Cup-to-disc ratio 0.0 (−0.3 to 0.4) 0.65 (<.001) 0.1 (−0.2 to 0.3) 0.77 (<.001)

AVG = average quadrant; INF = inferior quadrant; NAS = nasal quadrant; SUP = superior quadrant; RNFL = retinal nerve fiber layer; TEMP = temporal quadrant.

Agreement is expressed in terms of the mean difference between measurements from the 2 instruments (mean difference) and the 95% limits of agreements between the measurements (range). Correlation is expressed as the r 2 with the level of significance (P value). Mean difference = RTVue minus Stratus.