Table 1.

Angiographic parameters related with acute stent recoil assessment.

Parameters XINSORB (N = 8) EXCEL (N = 8) P value
Diameter of vessel segment chosen for deployment (mm) 2.78 ± 0.03 2.80 ± 0.03 0.18
Stent to artery ratio 1.09 ± 0.06 1.11 ± 0.02 0.47
Mean diameter of the inflated balloon (X, mm) 3.02 ± 0.13 3.06 ± 0.11 0.62
Mean lumen diameter of stent (Y, mm) 3.00 ± 0.20 3.14 ± 0.05 0.17
Mean lumen diameter of stent at 24 hours follow-up (mm) 3.00 ± 0.19 3.12 ± 0.08 0.23
Acute absolute recoil (X−Y, mm) 0.02 ± 0.13 −0.08 ± 0.08 0.19
Acute percent recoil ((X−Y)/X, %) 0.66 ± 4.32 −1.40 ± 3.83 0.45