Table 1.
Assessment of intra- and inter-observer variability of ascending and descending aortic measurements
|
INTRA-OBSERVER | ||||||
|---|---|---|---|---|---|---|
|
Location |
Ascending aorta (n = 15) |
Descending aorta (n = 15) |
||||
| Variable | c (m/s) | peak FCW (×10-5 m/s) | peak FEW (×10-5 m/s) | c (m/s) | peak FCW (×10-5 m/s) | peak FEW (×10-5 m/s) |
|
ICC |
0.961 |
0.894 |
0.391 |
0.976 |
0.869 |
0.88 |
|
95 % CI |
0.887-0.987 |
0.715-0.963 |
−0.153-0.754 |
0.929-0.992 |
0.654-0.984 |
0.680-0.958 |
|
Bias |
0.03 |
−0.2 |
−0.1 |
−0.1 |
0.3 |
0.2 |
|
Limits of agreement |
−0.7-0.76 |
−5.9-5.5 |
−2.9-2.6 |
−0.4-0.2 |
−6.6-7.2 |
−1.2-1.6 |
|
INTER-OBSERVER | ||||||
|
Location |
Ascending aorta (n = 15) |
Descending aorta (n = 15) |
||||
|
Variable |
c (m/s) |
peak FCW (×10-5 m/s) |
peak FEW (×10-5 m/s) |
c (m/s) |
peak FCW (×10-5 m/s) |
peak FEW (×10-5 m/s) |
|
ICC |
0.937 |
0.84 |
0.315 |
0.635 |
0.694 |
0.863 |
|
95 % CI |
0.903-0.989 |
0.588-0.943 |
−0.238-0.713 |
0.201-0.860 |
0.301-0.886 |
0.641-0.952 |
|
Bias |
−0.07 |
3.8 |
0.02 |
−0.2 |
4.2 |
0.3 |
| Limits of agreement | −0.96-0.83 | −2.1-9.6 | −1.8-1.9 | −1.5-1.1 | −5.1-13.5 | −1.2-1.9 |
ICC = intraclass correlation coefficient; CI = confidence interval; c = wave speed, FCW = forward compression wave; FEW = forward expansion wave.