Table 2. Organ dose (mGy) and effective dose (mSv) obtained in routine adult and paediatric chest CT examinations with 64-slice CT scanners.

CT scanner
Adult CT
Paediatric CT
A B C D A B C D
Tube voltage (kV) 120 120 120 120 120 120 120 120
The method of tube current adjustment Volume EC Smart mA Care Dose 4D D-DOM Volume EC Smart mA Care Dose 4D Fixed (203 mA)
Effective mAsa 186 134 266 112 37 28 116 100
Rotation time (s rotation–1) 0.5 0.4 0.5 0.5 0.5 0.4 0.33 0.5
Beam width (mm) 0.5 mm × 64 0.625 mm × 64 0.6 mm × 32 0.625 mm × 64 1.0 mm × 32 0.625 mm × 32 0.6 mm × 32 0.625 mm × 64
Pitch factor 0.828 1.375 1.0 1.015 1.406 1.375 1.0 1.015
Planned scan length (mm) 305 300 300 260 205 200 186 190
Exposed scan length (mm) 360 371 332 325 300 234 218 260
Tissue or organ Organ dose (mGy) Organ dose (mGy)
Brainb 0.3 0.1 0.5 0.6
Lensb 0.1 <0.1 0.3 0.3
Salivary glandsb 4.1 0.5 1.6 3.8
Thyroid 19.9 34.8 26.8 17.2 8.5 6.4 17.0 13.5
Lung 32.4 16.5 31.2 12.1 8.8 4.9 16.1 12.6
Breast 23.6 11.3 18.4 8.2 8.4 4.1 12.5 12.0
Oesophagus 29.1 15.7 28.5 10.8 8.3 3.6 14.3 12.1
Liver 27.7 9.7 19.7 10.7 7.7 4.1 12.3 12.2
Stomach 31.0 11.8 16.6 12.0 7.7 1.6 3.4 6.5
Colon 3.1 1.7 1.7 1.1 0.9 0.3 0.8 1.1
Ovaries 0.2 0.1 0.2 0.1 0.2 0.1 0.3 0.4
Bladder 0.1 0.1 0.1 <0.1 0.1 <0.1 0.2 0.2
Testes 0.1 <0.1 0.1 <0.1 0.1 <0.1 0.2 0.1
Bone surface 17.8 9.5 14.4 5.7 10.0 4.7 12.3 11.4
Red bone marrow 12.0 7.5 12.4 3.9 2.7 1.4 4.5 3.8
Skin 4.8 2.1 3.5 2.0 1.0 0.7 1.8 2.0
ICRP 103
Remainder tissues 17.6 10.1 14.1 7.7 6.3 3.1 7.5 7.5
Effective dose (mSv) 17.7 9.6 14.5 7.0 5.4 2.5 7.3 6.9
ICRP 60
Remainder tissues 15.9 7.6 11.8 5.8 5.5 2.3 5.1 5.5
Effective dose (mSv) 15.5 8.6 12.9 6.2 4.5 2.1 6.2 5.8
CTDIvol (mGy) 18.7 11.8 19.2 8.8 3.5 2.5 6.7 5.9
DLP (mGy cm) 672 436 635 286 104 59 146 153

CTDI, CT dose index; DLP, dose–length product; ICRP, International Commission on Radiological Protection.

a
graphic file with name bjr-82-1010-e001.jpg

bDoses not measured for the adult phantom were assumed to be zero.