Table 3. Organ dose (mGy) and effective dose (mSv) obtained in routine adult and paediatric abdominopelvic CT examinations with 64-slice CT scanners.

Adult CT
Paediatric CT
CT scanner A B C D A B C D
Tube voltage (kV) 120 120 120 120 120 120 120 120
The method of tube current adjustment Volume EC Smart mA Care Dose 4D D-DOM Volume EC Smart mA Care Dose 4D Fixed (203 mA)
Effective mAsa 208 124 251 247 43 28 90 100
Rotation time (s rotation–1) 0.5 0.5 0.5 0.5 0.5 0.5 0.33 0.5
Beam width (mm) 1.0 mm × 32 0.625 mm × 64 0.6 mm × 32 0.625 mm × 64 1.0 mm × 32 0.625 mm × 32 0.6 mm × 32 0.625 mm × 64
Pitch factor 0.844 0.984 1.0 0.641 1.406 1.375 1.0 1.015
Planned scan length (mm) 440 420 460 425 260 280 275 260
Exposed scan length (mm) 497 488 492 482 355 314 308 325
Tissue or organ Organ dose (mGy) Organ dose (mGy)
Brainb 0.1 <0.1 0.1 0.1
Lensb <0.1 <0.1 0.1 0.1
Salivary glandsb 0.1 <0.1 0.1 0.2
Thyroid 0.5 0.3 0.6 0.7 0.6 0.1 0.5 0.5
Lung 12.0 7.4 12.7 12.3 6.5 2.1 6.4 6.8
Breast 17.9 11.3 14.9 16.1 10.0 0.6 3.8 7.3
Oesophagus 13.6 5.9 12.6 13.3 6.1 1.8 6.3 6.4
Liver 27.1 11.5 21.9 24.9 8.5 4.5 12.2 12.2
Stomach 31.0 11.9 22.2 29.6 7.9 4.6 12.3 13.7
Colon 31.3 15.1 23.9 26.6 8.2 5.4 12.2 13.1
Ovaries 31.1 19.6 28.7 22.9 8.1 5.2 12.3 11.2
Bladder 32.6 17.3 27.5 20.1 8.6 5.2 11.3 11.3
Testes 21.0 6.5 7.2 10.3 6.5 3.2 3.7 8.9
Bone surface 29.4 17.8 23.4 21.2 10.1 4.4 9.7 8.2
Red bone marrow 14.4 8.2 12.6 11.8 4.1 1.8 4.8 3.9
Skin 5.8 4.0 5.0 6.1 1.0 1.2 2.0 2.5
ICRP 103
Remainder tissues 21.3 9.9 16.5 19.1 6.2 3.4 8.7 8.7
Effective dose (mSv) 20.7 10.3 16.6 17.8 6.8 3.0 7.8 8.5
ICRP 60
Remainder tissues 22.1 9.7 16.8 19.7 6.1 3.4 8.8 8.7
Effective dose (mSv) 21.9 10.7 17.2 18.0 6.8 3.3 8.1 8.9
CTDIvol (mGy) 22.9 10.3 18.1 17.6 4.2 2.2 5.2 5.9
DLP (mGy cm) 1138 503 891 848 148 69 160 192

CTDI, CT dose index; DLP, dose–length product; ICRP, International Commission on Radiological Protection.

a
graphic file with name bjr-82-1010-e002.jpg

bDoses not measured for the adult phantom were assumed to be zero.