Table 4. Doses for tissues and organs positioned within and at the boundaries of the scan length in adult and paediatric chest CT examinations with CT scanner A.

Adult CT Paediatric CT
Tube voltage (kV) 120 120 120 120 120 120 120 120
Rotation time (s rotation–1) 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5
Beam width (mm) 1.0 mm × 16 1.0 mm × 16 1.0 mm × 32 1.0 mm × 32 1.0 mm × 16 1.0 mm × 16 1.0 mm × 32 1.0 mm × 32
Pitch factor 0.938 1.438 0.844 1.406 0.938 1.438 0.844 1.406
Effective mAs 100 100 100 100 50 50 50 50
Planned scan length (mm) 305 305 305 305 205 205 205 205
Exposed scan length (mm) 340 356 364 400 240 256 264 300
Tissue or organ Organ dose (mGy) Organ dose (mGy)
Salivary glands 1.5 2.2 2.8 6.7
Thyroid 21.9 24.1 21.5 25.1 11.6 11.6 10.8 13.2
Lung 15.4 15.7 14.1 14.2 10.6 11.2 10.6 10.7
Breast 10.5 7.2 9.7 11.1 8.3 7.3 8.4 9.4
Oesophagus 14.5 14.4 13.6 14.3 10.7 10.8 10.2 10.3
Liver 12.5 12.5 12.3 13.2 9.6 10.5 10.0 10.2
Stomach 12.9 15.3 13.2 11.9 5.2 6.6 6.8 8.6
Colon 1.1 1.4 1.4 2.0 0.6 0.8 0.8 2.5