Table 5. Doses for tissues and organs positioned within and at the boundaries of the scan length in adult and paediatric abdominopelvic CT examinations with CT scanner A.

Adult CT Paediatric CT
Tube voltage (kV) 120 120 120 120 120 120 120 120
Rotation time (s rotation–1) 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5
Beam width (mm) 1.0 mm × 16 1.0 mm × 16 1.0 mm × 32 1.0 mm × 32 1.0 mm × 16 1.0 mm × 16 1.0 mm × 32 1.0 mm × 32
Pitch factor 0.938 1.438 0.844 1.406 0.938 1.438 0.844 1.406
Effective mAs 100 100 100 100 50 50 50 50
Planned scan length (mm) 440 440 440 440 260 260 260 260
Exposed scan length (mm) 475 491 499 535 295 311 319 355
Tissue or organ Organ dose (mGy) Organ dose (mGy)
Lung 5.9 6.7 6.8 7.7 5.4 6.1 5.4 6.3
Breast 3.1 10.1 8.1 10.4 1.9 2.5 5.2 6.6
Oesophagus 7.4 7.6 7.3 7.6 4.6 5.0 4.8 5.4
Liver 13.7 13.4 13.5 12.6 10.7 11.2 9.6 10.8
Stomach 16.0 13.5 14.5 16.9 11.3 11.8 10.3 9.9
Colon 15.5 14.9 14.4 13.8 11.2 11.3 10.5 11.0
Ovaries 15.8 16.1 14.7 12.8 9.1 9.2 8.7 8.9
Bladder 14.9 15.5 15.2 13.4 9.3 8.5 9.3 9.4
Testes 4.1 6.6 8.6 17.1 2.0 2.5 3.4 7.0