Table 1. Criteria for evaluation of satisfactory placement and tubal occlusion on hysterosalpingogram (HSG).

Placement
Satisfactory • Distal end of the inner coil being within the tube with less than half of the length of the inner coil trailing into the uterine cavity
• Proximal end of the inner coil is up to 30 mm into the tube from the cornua
Unsatisfactory • Absence of the microinsert on imaging (implying expulsion)
• More than half of the length of the inner coil trailing into the uterine cavity
• Proximal end of the microinsert in the tube more than 30 mm from the cornua
• Presence of the microinsert within the peritoneal cavity
Occlusion
Satisfactory (grade 1) • Complete non-filling of the tubes
Satisfactory (grade 2) • Filling of the tubes but not beyond the distal-most aspect of the outer coil
Tubal patency (grade 3) • Contrast beyond the outer coil
• Spill into the peritoneal cavity