Published in final edited form as: Nucl Instrum Methods Phys Res A. 2012 Nov 21;693:236–243. doi: 10.1016/j.nima.2012.07.029

Table I.

Summary of measured detector performance of all crystals. The timing of the 1.5×1.5×40 mm3 crystal is unreliable at the central depth region, and the 6 ns is a conservative estimate of the worst timing resolution for that crystal.

Crystal Properties DOI Resolution
(mm)
Energy Resolution
(%)
Timing Resolution
(ns)
Peak Shifting
(%)

Geometry
(mm)
Surface
Finish
Mean Max Min Mean Max Min Mean Max Min Mean Max
1.5×1.5×20 saw cut 1.5 2.0 1.2 17 19 15 2.6 2.9 2.1 15.4 26.4
1.5×1.5×20 30 µm 1.6 2.2 1.4 14 16 13 2.2 2.5 2.0 2.3 4.1
1.5×1.5×20 12 µm 2.6 3.0 2.2 14 21 13 2.4 2.6 2.2 2.2 5.8
1.5×1.5×20 9 µm 2.2 2.8 1.9 13 14 13 2.3 2.4 2.0 2.4 4.6
1.5×1.5×20 5 µm 3.3 4.4 2.1 14 16 13 2.4 2.6 2.2 2.9 7.0
1.5×1.5×20 0.5 µm 5.8 7.8 2.5 15 18 14 2.4 2.7 2.2 7.8 12.3
2.0×2.0×20 saw cut 1.6 2.1 1.2 14 16 13 2.3 2.5 2.1 2.1 3.8
2.0×2.0×20 30 µm 1.8 2.7 1.5 15 22 14 2.4 2.6 2.1 1.7 2.9
2.0×2.0×20 12 µm 2.8 4.1 2.4 13 16 12 2.2 2.4 2.1 2.3 4.7
2.0×2.0×20 9 µm 3.7 4.1 2.8 14 18 13 2.3 2.6 2.1 1.0 3.4
2.0×2.0×20 5 µm 5.9 7.3 3.9 15 16 14 2.4 2.7 2.2 4.3 6.0
2.0×2.0×20 0.5 µm 7.6 9.9 3.6 16 21 14 2.4 2.7 2.2 7.9 10.5
1.5×1.5×30 saw cut 2.4 3.9 1.9 18 21 15 3.3 5.2 2.2 23.8 40.1
1.5×1.5×40 saw cut 4.3 6.9 2.7 23 29 16 -- 6.0* 2.5 36.6 57.9
2.0×2.0×30 saw cut 2.8 3.9 2.3 16 17 14 2.6 2.8 2.2 11.1 19.1
2.0×2.0×40 saw cut 3.1 5.1 2.2 16 18 13 2.6 3.3 2.0 18.0 31.2