Published in final edited form as: Ultrason Imaging. 2012 Jul;34(3):159–180. doi: 10.1177/0161734612455576

Table 1.

Experimental Results (Copper Wire)

Medium Literature-Predicted Cwater, m/s Literature-Predicted Coil, m/s Local Estimate between Targets, m/s Error Relative to Cwater, m/s (%) Standard Deviation of Local Error, m/s (%)
With lens
Water only 1483.9 NA 1487.1 3.18 (0.21) 10.3 (0.69)
Castor oil (6 mm) on water 1489.1 1498.2 1476.2 −13.00 (−.88) 1.04 (0.071)
Castor oil (11 mm) on water 1492.4 1494.7 1478.8 −13.64 (−0.91) 1.19 (0.080)
Lens removed
Water only 1483.9 NA 1482.4 −1.44 (−0.097) 6.44 (0.43)
Castor oil (6 mm) on water 1489.1 1498.2 1491.5 2.27 (0.15) 0.89 (0.060)
Castor oil (11 mm) on water 1492.4 1494.7 1493.0 0.62 (0.041) 1.05 (0.070)

NA = Not applicable