Published in final edited form as: Phys Med Biol. 2012 Oct 3;57(21):6827–6848. doi: 10.1088/0031-9155/57/21/6827

Table 2.

The sampling parameters of the Focus-120 animal PET scanner’s system matrices used for image reconstruction reported in this work. Note that the LOR lengths between different detector blocks vary. The specific descriptions of the γγ, 18F and 15O sources are described in section 2.5. The size of the final system matrix is 64×64×19×5×21×9 or ~ 74 million elements. To facilitate visual presentation, a different set of parameters was used for the LOR-PDFs shown in Figures 5-8.

Variable Range Number of samples Sample size
r [−5,5]mm for γγ ,18F;
[−10,10]mm for 15O
64 ~0.16mm for γγ,18F;
~0.32mm for 15O
u [−5,5] mm for γγ, 18F;
[−10,10]mm for 15O
64 ~0.16mm for γγ, 18F;
~0.32mm for 15O
d [0, Lb] mm, Lb ≈ 40.7 to 157.2 19 ~2.1 to 8.3 mm
φ [φb-7.5°, φb+7.5° 5
b [2, 22] 21 1 block index
θ [−49.5°,49.5°] 9 11°