Skip to main content
. Author manuscript; available in PMC: 2013 Nov 1.
Published in final edited form as: J Magn Reson Imaging. 2012 Jul 23;36(5):1222–1230. doi: 10.1002/jmri.23761

Table 2.

The heart water and fat peak line-widths presented in Hertz (Hz) and ppm (part per million) as mean ± standard deviation - peaks at 1.3 ppm and 0.9 ppm for the ex-vivo data (N=70). A value of p less than 0.05 denotes significant difference in the 1.5-T and 3-T values.

Ex-vivo In-vivo

Water Methylene Methyl Water Methylene Methyl
4.7 ppm 1.3 ppm 0.9 ppm 4.7 ppm 1.3 ppm 0.9 ppm
3T (Hz) 19.6±14.1 29.5±9.7 34.5±28.7 23.1±6.0 26.2±10.3 24.1±18.7
1.5T (Hz) 10.5±6.4 16.7±5.8 25.8±17.4 17.8±8.2 17.8±6.6 12.4±9.2
p (1.5T vs. 3T) <0.05 <0.05 <0.05 <0.05 <0.05 <0.05

3T (ppm) 0.16±0.11 0.24±0.08 0.28±0.23 0.19±0.05 0.21±0.08 0.20±0.15
1.5T (ppm) 0.17±0.1 0.27±0.09 0.41±0.28 0.29±0.13 0.29±0.11 0.24±0.13
p (1.5T vs. 3T) 0.47 <0.05 <0.05 <0.05 <0.05 0.63