Table 2.
Reproducibility of CMR and Echocardiographic EF
CMR |
Echocardiography |
|||
---|---|---|---|---|
Intraobserver | Interobserver | Intraobserver | Interobserver | |
ICC (95% CI) |
0.98 (0.96 – 0.99) |
0.99 (0.98 – 0.99) |
0.94 (0.90 – 0.97) |
0.93 (0.88 – 0.96) |
Pearson’s Correlation Coefficient |
0.98 |
0.98 |
0.90 |
0.90 |
Bland-Altman Bias |
1.63 |
−1.1 |
1.8 |
−1.7 |
Limits of Agreement | −5.3, 8.5 | −7.0, 4.7 | −10.6, 14.3 | −13.7, 10.3 |