Skip to main content
. 2012 Sep 25;14(1):67. doi: 10.1186/1532-429X-14-67

Table 1.

MRI sequences for brain scan

 
Sequences for Philips MRI Scanner
  Body Region: Head, Patient Position: Supine, Patient Entry: Head First, Landmark on Eyes
 
3 plane localizer
Reference Scan
AX SE T1
AX FSE T2
T2 FLAIR
Diffusion
3D TOF MT
CE-MRA
Perfusion
 
Landmark on Eyes
Landmark on Eyes
Landmark on Eyes
Landmark on Eyes
Landmark on Eyes
Landmark on Eyes
Landmark on Eyes
Landmark on Eyes
Landmark on Eyes
Name Anatomy Philips
Head CNS-Brain
Ref Scans
Head CNS-Brain
Head CNS-Brain
Head CNS-Brain
Head CNS-Function
Head CNS-Angio
Head CNS-CE-Angio
Head CNS-Function
Sequence Name Philips
survey
RefScan
T1W_SE
T2W_TSE
T2W_FLAIR
DWI_E
S3DI_MC_HR
3D_DYN_AVM
sPRESTO
plane
ax/cor/sag
-
ax
ax
ax
ax
ax
sag
ax
Scan time
0:17
0:51
02:39
02:13
03:18
00:36
02:52
00:51
0:50
stacks
3
2
1
1
1
1
1
1
1
slices
3 each
82
18
18
16
18
100
10
30
Scan technique
T1-TFE
SENSE
SE
SE
IR
SE
FFE
FFE
FFE
Fast imaging mode
TFE
-
MS
TSE
TSE
EPI
none
none
EPI
(Turbo/EPI) Factor
42
-
1
19
23
89
-
-
19
TE [ms]
5.2
0.95
15
100
120
104
6.9
1.27
shortest
TR [ms]
15
8
483
3607
6000
3034
24
5
22
TI [ms]
-
-
-
-
2000
-
-
-
-
Flip angle [deg]
20
7
140
90
100
90
20
35
7
FOV [mm]
250
-
230
230
230
230
200
260
220
Scan [%]
74
82
79
74
68
79.6
57.4
100
55.8
Scan matrix
256 x 126
64 x 52
256 x 163
384 x 228
208 x 115
112 x 89
332 x 190
160 x 144
64 x 29
Slice thickness [mm]
acq/rec
10
-
5
5
6
5
1 / 0.5
16 / 8
3.5
Slice gap
0
 
1
1
2
1
-
-
-
NSA / NEX
1
3
2
3
2
1
1
1
1
Bandwidth [Hz/pixel] 186.9 2540.7 109.3 224.0 152.9 1834 108.8 541.1 60.3