The number of positive bacterial cultures obtained from rinse samples from biopsy channels of gastroscopes and colonoscopes and from swab samples obtained from automated endoscope reprocessors used for the decontamination of gastroscopes and colonoscopes

Category BC No. (%) AER No. (%) p value
Gastroscope, n = 300
32 (10.7)*
6 (2.0)
<0.0001
Colonoscope, n = 120
25 (20.8)*
1 (0.8)
<0.0001
Total, n = 420 57 (13.6) 7 (1.7) <0.0001

BC, biopsy channel; AER, automated endoscope reprocessor.

Pearson chi-squared test showed that the number of positive bacterial cultures from the samples obtained from BCs was significantly higher than that from the samples obtained from the AERs (p < 0.0001).