Abstract
Although high frequency ventilation (HFV) is an effective mode of ventilation, there is limited information available in regard to lung dynamics during HFV. To improve the knowledge of lung function during HFV we have developed a novel lung imaging and analysis technique. The technique can determine complex lung motion information in vivo with a temporal resolution capable of observing HFV dynamics. Using high-speed synchrotron based phase contrast X-ray imaging and cross-correlation analysis, this method is capable of recording data in more than 60 independent regions across a preterm rabbit lung in excess of 300 frames per second (fps). This technique is utilised to determine regional intra-breath lung mechanics of preterm rabbit pups during HFV. Whilst ventilated at fixed pressures, each animal was ventilated at frequencies of 1, 3, 5 and 10 Hz. A 50% decrease in delivered tidal volume was measured at 10 Hz compared to 1 Hz, yet at the higher frequency a 500% increase in minute activity was measured. Additionally, HFV induced greater homogeneity of lung expansion activity suggesting this ventilation strategy potentially minimizes tissue damage and improves gas mixing. The development of this technique permits greater insight and further research into lung mechanics and may have implications for the improvement of ventilation strategies used to support severe pulmonary trauma and disease.
Introduction
Conventional ventilation (CV) is commonly used to support breathing in both newborn and adult patients. If applied incorrectly, CV can cause ventilator induced lung injury (VILI), due to atelectasis (repeated opening and closing of alveoli) or overdistention of lung tissue [1], [2], [3]. Ventilation with smaller tidal volumes (Vt) has been shown to minimize lung damage [2], [3], [4], [5], [6]. As such, high frequency ventilation (HFV) may reduce VILI through the delivery of smaller volumes at higher ventilation rates, allowing increased minute volumes (product of tidal volume and frequency) and CO2 clearance [7], [8], [9]. Human [2], [5], [10], [11], [12] and animal [13], [14] studies indicate that HFV is an effective and safe mode of ventilation, however, there has been much inconsistency as to the specific HFV parameters that should be applied [15].
Typically during HFV inflations are delivered at 3 Hz to 15 Hz [16] using small tidal volumes that can potentially be less than the anatomical dead space [7], [17], [18]. Thus, the principal mechanism of gas exchange cannot be bulk gas transport, as occurs during normal respiration [19]. The underlying gas exchange mechanisms have been the subject of much debate [20], [21] and are not yet fully understood [19], [20], [22], [23]. It is proposed that increased minute volumes, along with enhanced gas mixing mechanisms, effectively and safely promote gas exchange during HFV [19], [20], [21], [24].
Although much research has focussed on optimizing HFV [25], [26], [27], [28], [29], [30], major improvements have been limited by a lack of knowledge of regional lung function during HFV. In particular, a regional understanding of tissue mechanics and gas transport is required to understand how the smaller respiratory units interact to effect efficient gas transfer [20], [21], [24]. Furthermore, the information must be obtained with sufficient temporal resolution to observe the dynamics within the respiratory cycle [31].
At the frequencies employed in HFV, imaging the lungs with sufficient temporal and spatial resolution is not possible with standard imaging approaches. Several techniques such as electrical impedance tomography (EIT), respiratory inductance plethysmography (RIP), magnetic resonance imaging (MRI) and X-ray computed tomography (CT) have been applied to investigate the lung during HFV. Whilst providing important information, each of these techniques possess specific limitations that restrict their ability to investigate lung dynamics during HFV. For instance, EIT [32] provides poor spatial resolution in addition to typically having temporal resolutions below 44 Hz [33], [34], [35], [36]. Although RIP can measure lung volume changes, it provides no spatial information on gas distribution within the lung [37], [38]. MRI and CT both offer higher spatial resolution than EIT [22], [39], [40], [41], [42], [43], but acquisition times at these higher spatial resolutions often require measurements to be made over multiple breath cycles [44], [45], [46] especially during HFOV conditions. Image blurring, due to a combination of lung motion and exposure times, has greatly limited the use of imaging to assess regional lung function with high spatial resolution [44].
Conventional (absorption based) X-ray imaging provides very poor levels of contrast in the lung. However, the lung (with its many tissue/air boundaries) is ideal for a method known as phase contrast X-ray imaging and for the lung this method provides images of high contrast and high detail [47]. Synchrotron X-ray sources provide highly coherent monochromatic X-rays that are well suited to phase contrast imaging. By combining this imaging method with velocimetry techniques, X-ray velocimetry was developed [48], [49]. X-ray velocimetry can non-invasively and accurately measure complex patterns of motion in opaque samples [48], [50], [51]. The application of X-ray velocimetry to the lungs results in vector fields defining the speed and direction of local lung tissue motion between consecutive frames, providing information on local lung mechanics with high spatial and temporal resolution [49], [52], [53]. Our aim was to regionally analyse the effect of ventilation frequency on lung tissue behaviour during HFV for the first time.
Materials and Methods
Ethics Statement
All animal procedures were approved by the SPring-8 Animal Care Committee and Monash University’s School of Biomedical Science’s Animal Ethics Committee. All studies were conducted in experimental hutch 3 of beamline 20B2, in the Biomedical Imaging Centre at the SPring-8 synchrotron in Japan.
Animal Preparations
Pregnant New Zealand white rabbits (preterm age of 28 days of gestation; term = 32d) were anaesthetised (propofol; i.v.; 12 mg/kg bolus), intubated and anaesthesia was maintained (isoflourane inhalation 1.5–4%). Pups were delivered by caesarean section, sedated (Nembutal; 0.1 mg, i.p.) and intubated with an endotracheal (ET) tube (18G) before the umbilical cord was cut. The pup was positioned upright in a warm water-filled plethysmograph (head out) located in the X-ray beam within the imaging hutch as previously described [54]. The ET tube was connected to the ventilator and animals were ventilated using a custom, pressure controlled small-animal ventilator [55]. Animals were ventilated using CV to fully aerate the lung and establish a functional residual capacity before HFV commenced. Airway pressure was oscillated between a positive end expiratory pressure (PEEP) of 9 cmH2O and a peak inspiratory pressure (PIP) of 21 cmH2O with ventilation frequencies of 1, 3, 5 and 10 Hz. Airway pressure and air flow were measured at the mouth opening using a pressure sensor and a pneumotach, respectively. Tidal volumes were calculated by integrating the flow signal.
Imaging Setup and Parameters
The highly coherent and bright X-ray source at beamline 20B2 offers the potential to perform in vivo studies with high spatial and temporal resolution. Radiation generated at the bending magnet was filtered to provide monochromatic X-rays (24 keV) and was directed through the pup in the water-filled plethysmograph (Figure 1). Images were acquired using a scintillator (Hamamatsu P43), image intensifier (Lambert II18 Gen3) and high-speed camera (IDT Y4; 1016×1016 pixel sensor). The propagation distance from the animal to the detector was 3 m. The combined effect of sensor pixel size and optical magnification yields an effective pixel size of 20.6 µm.
Image Analysis
All images were flat field corrected to improve image quality [56]. A procedure to remove low frequency artefacts, such as intensity gradients due to the use of a cylindrical water-filled plethysmograph, was also applied. In this procedure all rows in the image were scaled to have the same average pixel intensity, which was then successively applied to scaling of the columns, and repeated.
The high contrast lung speckle, as seen in Figure 2A, Video S1, Video S2, provides a rich level of detail for cross-correlation analysis of motion at high spatial resolution [53]. Bone structure within the images was minimised via a band-pass spatial frequency filter to improve image quality and minimise errors in the cross-correlation analysis of the lung motion. A mask image was created which constrained the X-ray velocimetry measurements to regions of lung tissue.
X-ray Velocimetry Analysis
A cross-correlation analysis is performed on the images by first dividing the image into regions, known as interrogation windows. An FFT-based cross-correlation is then performed on each interrogation window between subsequent frames. The peak of the resulting cross-correlation is the modal value for displacement within that particular window at that point in time. Window sizes are chosen to improve resolution whilst minimising errors. Large windows include more lung speckle thus providing improved signal to noise ratio (SNR). Smaller windows provide better resolution at the cost of the SNR. Furthermore, too small a window and the speckle may move outside of the window, thus creating additional errors. Noisier images require larger windows to get a statistically significant answer, therefore the spatial resolution decreases with an increase in image noise via the larger windows required for analysis. For a more in depth understanding of the process refer to [57].
To determine any motion of the animal as a whole a cross-correlation analysis was performed using large interrogation windows that included the bone structure of the animal; this was subtracted from the motion of the lungs. The X-ray velocimetry measurements were performed using in-house software as used in [52] and as described in detail in [57] with a detailed error analysis in [58]. The cross-correlation utilised an interrogation window size of 64×64 pixels and 75% window overlap. An iterative approach was applied with 128×128 pixel windows used first to get an approximation of the local motion; this information was then used in the subsequent 64×64 analysis to increase dynamic range and spatial resolution (see for example [48]). The analysis provided a vector spacing of 330 µm resulting in over 1000 measurements in total across the lung at each time point. The result can be seen in Figure 2B as a vector map.
Instantaneous local lung tissue expansion has been shown to be derived from the spatial derivatives of the velocity field [52]. Figure 2C shows a map of the local lung tissue expansion between two successive frames. Expansion activity is defined as the root-mean-square (RMS) of the local tissue expansion through time. Expansion activity includes the degree of lung expansion activity during both inspiration and expiration phases. The 2D X-ray velocimetry measurements at each location are a representation of the motion of lung tissue through the entire depth of the lung at that particular location.
Local tissue expansion was integrated across the lung to give instantaneous respiratory flow. Respiratory flow was integrated through time to give respiratory volume. Respiratory volume was calibrated using the method described in [52] to determine the tidal volume in millilitres.
Results and Discussion
Figure 3A shows respiratory volume measured using our imaging technique for a ventilation rate of 1 Hz. This measurement was found to faithfully reproduce the tidal volume measured via a pnuemotach (coefficient of determination >0.96). Figure 3B presents maps of accumulative regional lung expansion calculated with respect to the onset of inspiration at three time points that are labelled in Figure 3A. Each data point on the respiratory volume trace in Figure 3A is the sum of more than 1000 locations across the lung, as seen in the expansion maps in Figure 3B. This data clearly demonstrates the spatial and temporal resolution that can be achieved using this technique and the potential for using it to assess how different ventilation strategies alter the distribution of ventilation within the lung.
Increasing inflation frequency whilst maintaining fixed end-inspiration and end-expiration pressures reduced the magnitude of expansion activity (Figure 4A). Expansion activity was normalised to that delivered at 1 Hz to account for inter-animal variability. There was an average decrease in expansion activity of 50% at 10 Hz compared to 1 Hz. The frequency dependent decrease in expansion activity is due to the decrease in inspiration time which limits the duration over which gas can flow into the lung, as previously described [59]. A decrease in expansion activity is representative of a decrease in tidal volume which is thought to be a necessary and an advantageous outcome of higher ventilation frequencies [2].
An important finding of this study is that with increasing frequency the magnitude of lung tissue expansion activity throughout the lung markedly decreased (Figure 4B), simultaneously with increasing uniformity of ventilation throughout the lung (Figure 5). These findings demonstrate that at higher frequencies the lung behaves more uniformly which likely protects areas of lung tissue from over-distension. When ventilated at lower frequencies, lung tissue has time to expand according to its local compliance. However at rates of ventilation that are faster than the time constant of the lung tissue, the effect of localised differences in tissue compliance is minimised resulting in more uniform inflation.
The minute activity, calculated as the expansion activity multiplied by the number of cycles per minute, was found to increase as the rate of ventilation increased (Figure 6) despite decreased tidal volumes. At 10 Hz, there was approximately a 5-fold increase in minute activity compared to 1 Hz. Therefore, over the same duration, more gas would be exchanged with the animal at the higher frequencies than would at the lower frequencies, despite the use of smaller tidal volumes. This high amount of gas transport activity allows for several of the postulated HFV gas exchange mechanisms [20], [21], [24] to exist. The tidal volumes delivered varied from ∼0.8 ml at 1 Hz to ∼0.4 ml at 10 Hz. Siew et al. [60] have shown that for animals of this breed and weight the respiratory dead space is approximately 0.08 ml. Therefore the total effect respiratory dead space would not change the trends observed. Despite the use of shorter ventilation cycles, HFV was capable of providing fresh gas to the animals with each cycle, whilst minimizing lung tissue excursion per cycle. Further evidence of this can be seen in Video S1 and Video S2.
Current limiting factors to be overcome are mostly based around the translation of this imaging technology towards a laboratory or clinical environment. However, with a steady advance in X-ray source technology for example the advent of the liquid metal jet anode source [61], translation of this imaging method out of the synchrotron environment appears to be plausible.
Conclusion
Previously, global measures were the only methods with suitable temporal resolution for investigating HFV. The use of this 2D X-ray velocimetry technique allows for a regional and intra-breath analysis of HFV in vivo. We hypothesize that during HFV the increase in minute lung activity with a decrease in tidal volume would result in an increase in gas mixing and gas exchange whilst potentially minimizing lung damage by preserving uniformity of lung tissue expansion. This in vivo imaging and analysis technique provides capabilities not previously possible, including the measurement of in vivo lung function at rates of 300 fps with a spatial resolution of 330 um. This work provides greater knowledge of lung function during high frequency ventilation allowing for further research into regional lung dynamics.
Supporting Information
Funding Statement
Research is funded by the Australian Research Council (DP110101498), the National Health and Medical Research Council (491103), Victorian Governments Operational Infrastructure Support Program and supported by beamtime grants from the Japan Synchrotron Radiation Research Institute. We acknowledge travel funding provided by the International Synchrotron Access Program (ISAP) managed by the Australian Synchrotron and funded by the Australian Government. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1. Krishnan JA (2000) High-Frequency Ventilation for Acute Lung Injury and ARDS. Chest 118: 795–807. [DOI] [PubMed] [Google Scholar]
- 2. Brower RG, Matthay MA, Morris A, Schoenfeld D, Thompson BT, et al. (2000) Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. New England Journal of Medicine 342: 1301–1308. [DOI] [PubMed] [Google Scholar]
- 3. Seah ASSAS, Grant KA, Aliyeva M, Allen GB, Bates JHT (2011) Quantifying the Roles of Tidal Volume and PEEP in the Pathogenesis of Ventilator-Induced Lung Injury. Annals of Biomedical Engineering 39: 1505–1516. [DOI] [PubMed] [Google Scholar]
- 4. Carney D, DiRocco J, Nieman G (2005) Dynamic alveolar mechanics and ventilator-induced lung injury. Critical Care Medicine 33: S122–128. [DOI] [PubMed] [Google Scholar]
- 5. Amato MBP, Barbas CSV, Medeiros DM, Magaldi RB, Schettino GDP, et al. (1998) Effect of a protective-ventilation strategy on mortality in the acute respiratory distress syndrome. New England Journal of Medicine 338: 347–354. [DOI] [PubMed] [Google Scholar]
- 6. Clark RH (1994) High-Frequency Ventilation. Journal of Pediatrics 124: 661–670. [DOI] [PubMed] [Google Scholar]
- 7. Bohn DJ, Miyasaka K, Marchak BE, Thompson WK, Froese AB, et al. (1980) Ventilation by High-Frequency Oscillation. Journal of Applied Physiology 48: 710–716. [DOI] [PubMed] [Google Scholar]
- 8. Smith BE (1990) High-Frequency Ventilation - Past, Present and Future. British Journal of Anaesthesia 65: 130–138. [DOI] [PubMed] [Google Scholar]
- 9. Slutsky AS, Kamm RD, Rossing TH, Loring SH, Lehr J, et al. (1981) Effects Of Frequency, Tidal Volume, And Lung-Volume On Co2 Elimination In Dogs By High-Frequency (2–30 Hz), Low Tidal Volume Ventilation. Journal of Clinical Investigation 68: 1475–1484. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Cools F, Askie LM, Offringa M, Asselin JM, Calvert SA, et al. (2010) Elective high-frequency oscillatory versus conventional ventilation in preterm infants: a systematic review and meta-analysis of individual patients’ data. Lancet 375: 2082–2091. [DOI] [PubMed] [Google Scholar]
- 11. Wunsch H, Mapstone J (2005) High-frequency ventilation versus conventional ventilation for the treatment of acute lung injury and acute respiratory distress syndrome: A systematic review and cochrane analysis. Anesthesia and Analgesia 100: 1765–1772. [DOI] [PubMed] [Google Scholar]
- 12. Rimensberger PC, Beghetti M, Hanquinet S, Berner M (2000) First Intention High-Frequency Oscillation With Early Lung Volume Optimization Improves Pulmonary Outcome in Very Low Birth Weight Infants With Respiratory Distress Syndrome. Pediatrics 105: 1202–1208. [DOI] [PubMed] [Google Scholar]
- 13. Imai Y, Nakagawa S, Ito Y, Kawano T, Slutsky AS, et al. (2001) Comparison of lung protection strategies using conventional and high-frequency oscillatory ventilation. Journal of Applied Physiology 91: 1836–1844. [DOI] [PubMed] [Google Scholar]
- 14. Rimensberger PC, Pristine G, Mullen JBM, Cox PN, Slutsky AS (1999) Lung recruitment during small tidal volume ventilation allows minimal positive end-expiratory pressure without augmenting lung injury. Critical Care Medicine 27: 1940–1945. [DOI] [PubMed] [Google Scholar]
- 15. van Kaam AH, Rimensberger PC (2007) Lung-protective ventilation strategies in neonatology: What do we know - What do we need to know? Critical Care Medicine 35: 925–931. [DOI] [PubMed] [Google Scholar]
- 16.Sud S, Sud M, Friedrich JO, Meade MO, Ferguson ND, et al.. (2010) High frequency oscillation in patients with acute lung injury and acute respiratory distress syndrome (ARDS): systematic review and meta-analysis. British Medical Journal 340. [DOI] [PubMed]
- 17. Butler WJ, Bohn DJ, Bryan AC, Froese AB (1980) Ventilation by High-Frequency Oscillation in Humans. Anesthesia and Analgesia 59: 577–584. [PubMed] [Google Scholar]
- 18. Slutsky AS, Drazen FM, Ingram Jr RH, Kamm RD, Shapiro AH, et al. (1980) Effective pulmonary ventilation with small-volume oscillations at high frequency. Science 209: 609–609. [DOI] [PubMed] [Google Scholar]
- 19. Shabtai Y, Gavriely N (1989) Frequency And Amplitude Effects During High-Frequency Vibration Ventilation In Dogs. Journal of Applied Physiology 66: 1127–1135. [DOI] [PubMed] [Google Scholar]
- 20. Chang HK (1984) Mechanisms of gas transport during ventilation by high-frequency oscillation. Journal of Applied Physiology 56: 553–553. [DOI] [PubMed] [Google Scholar]
- 21. Pillow JJ (2005) High-frequency oscillatory ventilation: mechanisms of gas exchange and lung mechanics. Critical Care Medicine 33: S135–141. [DOI] [PubMed] [Google Scholar]
- 22. Terekhov M, Rivoire J, Scholz A, Wolf U, Karpuk S, et al. (2010) Measurement of Gas Transport Kinetics in High-Frequency Oscillatory Ventilation (HFOV) of the Lung Using Hyperpolarized (3)He Magnetic Resonance Imaging. Journal of Magnetic Resonance Imaging 32: 887–894. [DOI] [PubMed] [Google Scholar]
- 23. Heraty KB, Laffey JG, Quinlan NJ (2008) Fluid Dynamics of Gas Exchange in High-Frequency Oscillatory Ventilation: In Vitro Investigations in Idealized and Anatomically Realistic Airway Bifurcation Models. Annals of Biomedical Engineering 36: 1856–1869. [DOI] [PubMed] [Google Scholar]
- 24. Slutsky A, Drazen JM (2002) Ventilation With Small Tidal Volumes. New England Journal of Medicine 347: 630–631. [DOI] [PubMed] [Google Scholar]
- 25. Pellicano A, Tingay DG, Mills JF, Fasulakis S, Morley CJ, et al. (2009) Comparison of four methods of lung volume recruitment during high frequency oscillatory ventilation. Intensive Care Medicine 35: 1990–1998. [DOI] [PubMed] [Google Scholar]
- 26. Hager DN, Fessler HE, Kaczka DW, Shanholtz CB, Fuld MK, et al. (2007) Tidal volume delivery during high-frequency oscillatory ventilation in adults with acute respiratory distress syndrome. Critical Care Medicine 35: 1522–1529. [DOI] [PubMed] [Google Scholar]
- 27. Wood B, Karna P, Adams A (2002) Specific compliance and gas exchange during high-frequency oscillatory ventilation. Critical Care Medicine 30: 1523–1527. [DOI] [PubMed] [Google Scholar]
- 28. Habib RH, Pyon KH, Courtney SE (2002) Optimal high-frequency oscillatory ventilation settings by nonlinear lung mechanics analysis. American Journal of Respiratory and Critical Care Medicine 166: 950–953. [DOI] [PubMed] [Google Scholar]
- 29. Pillow JJ, Neil H, Wilkinson MH, Ramsden CA (1999) Effect of I/E ratio on mean alveolar pressure during high-frequency oscillatory ventilation. Journal of Applied Physiology 87: 407–414. [DOI] [PubMed] [Google Scholar]
- 30. Thome U, Pohlandt F (1998) Effect of the TI/TE ratio on mean intratracheal pressure in high-frequency oscillatory ventilation. Journal of Applied Physiology 84: 1520–1527. [DOI] [PubMed] [Google Scholar]
- 31. Bayat S, Porra L, Suhonen H, Janosi T, Strengell S, et al. (2008) Imaging of lung function using synchrotron radiation computed tomography: What’s new? European Journal of Radiology 68: S78–S83. [DOI] [PubMed] [Google Scholar]
- 32. Cheney M, Isaacson D, Newell JC (1999) Electrical impedance tomography. Siam Review 41: 85–101. [Google Scholar]
- 33. Wolf GK, Grychtol B, Boyd TK, Zurakowski D, Arnold JH (2010) Regional overdistension identified with electrical impedance tomography in the perflubron-treated lung. Physiological Measurement 31: S85–S95. [DOI] [PubMed] [Google Scholar]
- 34. van Genderingen HR, van Vught AJ, Jansen JRC (2004) Regional lung volume during high-frequency oscillatory ventilation by electrical impedance tomography. Critical Care Medicine 32: 787–794. [DOI] [PubMed] [Google Scholar]
- 35. Dunlop S, Hough J, Riedel T, Fraser JF, Dunster K, et al. (2006) Electrical impedance tomography in extremely prematurely born infants and during high frequency oscillatory ventilation analyzed in the frequency domain. Physiological Measurement 27: 1151–1165. [DOI] [PubMed] [Google Scholar]
- 36. Frerichs I, Schmitz G, Pulletz S, Schadler D, Zick G, et al. (2007) Reproducibility of regional lung ventilation distribution determined by electrical impedance tomography during mechanical ventilation. Physiological Measurement 28: S261–S267. [DOI] [PubMed] [Google Scholar]
- 37. Brazelton TB, Watson KF, Murphy M, Al-Khadra E, Thompson JE, et al. (2001) Identification of optimal lung volume during high-frequency oscillatory ventilation using respiratory inductive plethysmography. Critical Care Medicine 29: 2349–2359. [DOI] [PubMed] [Google Scholar]
- 38. Weber K, Courtney SE, Pyon KH, Chang GY, Pandit PB, et al. (2000) Detecting lung overdistention in newborns treated with high-frequency oscillatory ventilation. Journal of Applied Physiology 89: 364–372. [DOI] [PubMed] [Google Scholar]
- 39.van Beek EJR, Hoffman EA (2008) Functional imaging: CT and MRI. Clinics in Chest Medicine 29: 195−+. [DOI] [PMC free article] [PubMed]
- 40. Hoffman EA, Clough AV, Christensen GE, Lin CL, McLennan G, et al. (2004) The comprehensive imaging-based analysis of the lung: A forum for team science. Academic Radiology 11: 1370–1380. [DOI] [PubMed] [Google Scholar]
- 41. Tustison NJ, Cook TS, Song G, Gee JC (2011) Pulmonary Kinematics From Image Data: A Review. Academic Radiology 18: 402–417. [DOI] [PubMed] [Google Scholar]
- 42. Wolf U, Scholz A, Terekhov M, Koebrich R, David M, et al. (2010) Visualization of Inert Gas Wash-Out During High-Frequency Oscillatory Ventilation Using Fluorine-19 MRI. Magnetic Resonance in Medicine 64: 1479–1483. [DOI] [PubMed] [Google Scholar]
- 43. Biederer J, Hintze C, Fabel M, Dinkel J (2010) Magnetic Resonance Imaging and Computed Tomography of Respiratory Mechanics. Journal of Magnetic Resonance Imaging 32: 1388–1397. [DOI] [PubMed] [Google Scholar]
- 44. Easley RB, Lancaster CT, Fuld MK, Custera JW, Hager DN, et al. (2009) Total and regional lung volume changes during high-frequency oscillatory ventilation (HFOV) of the normal lung. Respiratory Physiology & Neurobiology 165: 54–60. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45. Mulreany DG, Simon BA, Murphy KJ, Easley RB (2009) Volumetric Xenon-CT Imaging of Conventional and High-frequency Oscillatory Ventilation. Academic Radiology 16: 718–725. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46. Luecke T, Herrmann P, Kraincuk P, Pelosi P (2005) Computed tomography scan assessment of lung volume and recruitment during high-frequency oscillatory ventilation. Critical Care Medicine 33: S155–162. [DOI] [PubMed] [Google Scholar]
- 47. Lewis RA, Yagi N, Kitchen MJ, Morgan MJ, Paganin D, et al. (2005) Dynamic imaging of the lungs using x-ray phase contrast. Physics in Medicine and Biology 50: 5031–5040. [DOI] [PubMed] [Google Scholar]
- 48.Fouras A, Dusting J, Lewis R, Hourigan K (2007) Three-dimensional synchrotron x-ray particle image velocimetry. Journal of Applied Physics 102.
- 49. Fouras A, Dusting J, Sheridan J, Kawahashi M, Hirahara H, et al. (2009) Engineering Imaging: Using Particle Image Velocimetry To See Physiology In A New Light. Clinical and Experimental Pharmacology and Physiology 36: 238–247. [DOI] [PubMed] [Google Scholar]
- 50. Lee SJ, Kim GB (2003) X-ray particle image velocimetry for measuring quantitative flow information inside opaque objects. Journal of Applied Physics 94: 3620–3623. [Google Scholar]
- 51. Jamison RA, Dubsky S, Siu KKW, Hourigan K, Fouras A (2011) X-ray Velocimetry and Haemodynamic Forces Within a Stenosed Femoral Model at Physiological Flow Rates. Annals of Biomedical Engineering 39: 1643–1653. [DOI] [PubMed] [Google Scholar]
- 52. Fouras A, Allison BJ, Kitchen MJ, Dubsky S, Nguyen J, et al. (2012) Altered Lung Motion is a Sensitive Indicator of Regional Lung Disease. Annals of Biomedical Engineering 40: 1160–1169. [DOI] [PubMed] [Google Scholar]
- 53.Fouras A, Kitchen MJ, Dubsky S, Lewis RA, Hooper SB, et al.. (2009) The past, present, and future of x-ray technology for in vivo imaging of function and form. Journal of Applied Physics 105.
- 54. Hooper SB, Kitchen MJ, Wallace MJ, Yagi N, Uesugi K, et al. (2007) Imaging lung aeration and lung liquid clearance at birth. FASEB 21: 3329–3337. [DOI] [PubMed] [Google Scholar]
- 55. Kitchen MJ, Habib A, Fouras A, Dubsky S, Lewis RA, et al. (2010) A new design for high stability pressure-controlled ventilation for small animal lung imaging. Journal of Instrumentation 5: T02002. [Google Scholar]
- 56. Kitchen MJ, Lewis RA, Morgan MJ, Wallace MJ, Siew ML, et al. (2008) Dynamic measures of regional lung air volume using phase contrast x-ray imaging. Physics in Medicine and Biology 53: 6065–6077. [DOI] [PubMed] [Google Scholar]
- 57. Fouras A, Lo Jacono D, Hourigan K (2008) Target-free Stereo PIV: a novel technique with inherent error estimation and improved accuracy. Experiments in Fluids 44: 317–329. [Google Scholar]
- 58. Fouras A, Dusting J, Hourigan K (2007) A simple calibration technique for stereoscopic particle image velocimetry. Experiments in Fluids 42: 799–810. [Google Scholar]
- 59. Scalfaro P, Pillow JJ, Sly PD, Cotting J (2001) Reliable tidal volume estimates at the airway opening with an infant monitor during high-frequency oscillatory ventilation. Critical Care Medicine 29: 1925–1930. [DOI] [PubMed] [Google Scholar]
- 60. Siew ML, te Pas AB, Wallace MJ, Kitchen MJ, Lewis RA, et al. (2009) Positive end-expiratory pressure enhances development of a functional residual capacity in preterm rabbits ventilated from birth. Journal of Applied Physiology 106: 1487–1493. [DOI] [PubMed] [Google Scholar]
- 61. Hemberg O, Otendal M, Hertz HM (2003) Liquid-metal-jet anode electron-impact x-ray source. Applied Physics Letters 83: 1483–1485. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.