Current PET/MR systems

PET/MR system Biograph mMR (Siemens) Ingenuity TF (Philips) Discovery PET/CT 690 + MR 750 (GE)
PET-system      
  Crystal material LSO LYSO LYSO
  Crystal elements dimension 4x4x20 mm3 4x4x22 mm3 4.7x6.3x25 mm3
  Photomultipliers No, 4032 APDs 420 1024
  Ring diameter 65.6 cm 90.3 cm 88.6 cm
  Transaxial FoV 59.4 cm 67.6 cm 70 cm
  Axial FoV 25.8 cm 18.0 cm 15.7 cm
  Energy window 430-610 keV 460-665 keV 425-650 keV
  Coincidence window 5.9 ns 6 ns 4.9 ns
  Time-of-flight No Yes Yes
MR-system      
  Field strength 3 Tesla 3 Tesla 3 Tesla
  Bore diameter 60 cm 60 cm 60 cm
  Max FoV 50x50x45 cm3 50x50x45 cm3 48x48x48 cm3
  Field homogeneity (40cm3) 0.25 ppm 0.5 ppm 0.25 ppm
PET/MR      
  Acquisition Simultaneous Sequential (same room) Sequential (2 separate rooms)
  PET-attenuation corr. MR-based MR-based CT-based