Skip to main content
. 2012 Nov;50(11):3747–3750. doi: 10.1128/JCM.02037-12

Table 3.

Differences in log2 dilutions of MICs obtained by Vitek2, Etest, and MTS compared to BMD

Test method No. (%) of isolates showing a MIC difference (in log2 dilutions) of:
>−3 −3 −2 −1 0 1 2 3
Vitek2 4 (1.7) 45 (18.7) 99 (41.1) 89 (36.5) 4 (1.7)
Etest 1 (0.4) 4 (1.7) 37 (15.4) 115 (47.7) 77 (32.0) 7 (2.9)
MTS 5 (2.1) 6 (2.5) 43 (17.8) 121 (50.2) 56 (23.2) 10 (4.1)