Table 4. Technical factors related to the DAP (Gy cm−2) at each fluoroscopy sequence delivered during 100 percutaneous transluminal coronary angioplasties performed by 3 practitioners.

Factors n Estimations on log(dose) scalea
Back-transformed coefficients p-value
Coefficients 95% CI
Constant 1550 1.35 (0.15, 2.56)b 3.86
Cinematography settings <0.0001
 “Low dose” mode at 12.5 fr s–1 1004 0 reference
 “Low dose” mode at 25 fr s–1 154 0.69 (0.40, 1.00)b 1.99
 “High dose” mode at 12.5 fr s–1 305 0.38 (0.19, 0.57)b 1.46
 “High dose” mode at 25 fr s–1 87 0.77 (0.46, 1.08)b 2.16
Focus–intensifier distance (cm) 1550 −0.00 (−0.01, 0.01) −1.00 0.30
Intensifier field size (cm) <0.0001
 14 131 0 reference
 17 1401 0.43 (0.23, 0.62)b 1.53
 23 18 0.45 (−0.01, 0.91) 1.57
Collimation <0.0001
 No 545 0 reference
 Yes 1005 −0.70 (−0.82, −0.58)b −2.01
X-ray tube orientation (vs mean effect of beam orientation) <0.0001
 CAUD/RAO (25–55)°C 84 0.29 (0.07, 0.51)b 1.34
 CAUD/RAO (15-25)°C 80 −0.15 (−0.37, 0.06) −1.17
 CAUD/RAO (5-15)°C 81 −0.07 (−0.30, 0.15) −1.07
 CAUD/PA 112 0.11 (−0.09, 0.32) 1.12
 CAUD/LAO (5-25)°C 62 −0.00 (−0.33, 0.33) −1.00
 CAUD/LAO (25–35)°C 49 0.21 (−0.08, 0.50) 1.64
 CAUD/LAO (35–100)°C 115 0.31 (0.13, 0.49)b 1.18
 PA/RAO (5–55)°C 34 −0.53 (−0.85, −0.22)b −1.70
 PA/PA 30 −0.12 (−0.53, 0.29) −1.13
 PA/LAO (5-25)°C 47 −0.18 (−0.54, 0.18) −1.20
 PA/LAO (25–35)°C 108 −0.09 (−0.31, 0.14) −1.09
 PA/LAO (35–55)°C 45 −0.05 (−0.34, 0.25) −1.05
 PA/LAO (55–100)°C 117 −0.17 (−0.37, 0.02) −1.20
 CRAN/RAO (35–55)°C 60 −0.21 (−0.47, 0.05) −1.24
 CRAN/RAO (25–35)°C 70 0.06 (−0.20, 0.33) 1.07
 CRAN/RAO (15-25)°C 104 0.05 (−0.16, 0.26) 1.05
 CRAN (35–45)°C/RAO (5-15)°C 72 0.16 (−0.09, 0.42) 1.18
 CRAN (5–35)°C/RAO (5-15)°C 65 −0.18 (−0.48, 0.12) −1.20
 CRAN/PA 72 0.01 (−0.22, 0.24) 1.01
 CRAN/LAO (5-25)°C 68 −0.16 (−0.43, 0.12) −1.17
 CRAN/LAO (25–35)°C 37 0.50 (0.16, 0.83)b 1.64
 CRAN/LAO (35–100)°C 38 0.17 (−0.14, 0.48) 1.18
Sequence 1550 −0.01 (−0.01, 0.00) −1.01 0.08

95% CI, confidence interval; CAUD, caudal; CRAN, cranial; fr, frames; LAO, left anterior oblique; PA, posters-anterior; RAO, right anterior oblique.

aBoxox transformed coefficients (λ=0).

bConfidence interval that does not include 0. Maximum likelihood estimation criterion; standard errors in the random part are 0.59 for the practitioners intercept, 0.41 for the patients intercept and 0.83 for the residual error. Residual error variance of a previous null model was 0.79. The residual variance proportion explained by the covariates is 13.2%; the practitioners' random intercept variance decreased by 36.9% while introducing the covariates.