Skip to main content
. 2012 Apr 11;13:56. doi: 10.1186/1471-2474-13-56

Table 3.

Results of correlation analysis

ASR Resurfacing        
Follow-up: 1 year
Cr
Co
Inclination
Arc of cover
Co
Spearmen: ,855; p = 0,007
 
Spearmen: -,229; p = 0,586
Spearmen: ,229; p = 0,586
Cr
 
Spearmen: ,855; p = 0,007
Spearmen: -,048; p = 0,911
Spearmen: ,048; p = 0,911
Follow-up: 2 years
Cr
Co
Inclination
Arc of cover
Co
Spearmen: ,695; p = 0,056
 
Spearmen: -,530; p = 0,177
Spearmen: ,539; p = 0,168
Cr
 
Spearmen: ,695; p = 0,056
Spearmen: -,778; p = 0,023
Spearmen: ,476; p = 0,233
ASR XL Head
 
 
 
 
Follow-up: 1 year
Cr
Co
Inclination
Arc of cover
Co
Spearmen: ,819; p < 0,001
 
Spearmen: ,192; p = 0,357
Spearmen: -,176; p = 0,389
Cr
 
Spearmen: ,819; p < 0,001
Spearmen: ,103; p = 0,623
Spearmen: -,107; p = 0,601
Follow-up: 2 years
Cr
Co
Inclination
Arc of cover
Co
Spearmen: ,805; p < 0,001
 
Spearmen: ,240; p = 0,201
Spearmen: -,326; p = 0,073
Cr   Spearmen: ,805; p < 0,001 Spearmen: ,234; p = 0,213 Spearmen: -,258; p = 0,161

Spearmen's correlation coefficient was calculated to determine correlations between metal ion concentration, inclination and arc of cover at, one and two years postoperatively.