Skip to main content
. 2012 Apr 20;25(6):720–728. doi: 10.1007/s10278-012-9474-6

Table 13.

Measurement values of the monitor’s uniformity depending on the difference in lifetime

Lifetime Center luminance Top left luminance Top right luminance Bottom left luminance Bottom right luminance p
Minimum value (D) 3,602 306.19 300.22 (−1.9 %) 302.04 (−1.4 %) 302.08 (−1.3 %) 304.93 (−0.4 %) 0.025*
Maximum value (C) 4,965 303.63 316.09 (4.1 %) 314.26 (3.5 %) 315.50 (3.9 %) 316.16 (4.1 %) 0.040*

Interaction effect using ANOVA test model and Duncan test model. Unit is the number and in candela per square meter

*p < 0.05