Table 1.

The Analytical precision of the method

Ceramides, µmol/L

C14:0 C16:0 C18:1 C18:0 C20:0 C24:0






Mean ± SD CV, % Mean ± SD CV, % Mean ± SD CV, % Mean ± SD CV, % Mean ± SD CV, % Mean ± SD CV, %
Intraassay precision
Low Conc.(n = 10) 0.26 ± 0.01 2.7 0.97 ± 0.02 1.5 0.31 ± 0.01 2.6 0.39 ± 0.01 2.5 0.40 ± 0.01 2 0.38 ± 0.01 2.6
High Conc. (n = 10) 2.61 ± 0.03 1.1 9.65 ± 0.10 1.1 3.37 ± 0.08 2.2 3.89 ± 0.05 1.4 4.04 ± 0.05 1.3 3.89 ± 0.05 1.3
Interassay precision
Low Conc.(n = 10) 0.26 ± 0.01 3.6 0.99 ± 0.03 3.3 0.33 ± 0.02 5.7 0.41 ± 0.02 4.1 0.42 ± 0.02 5.2 0.39 ± 0.02 5.1
High Conc. (n = 10) 2.69 ± 0.08 3 9.88 ± 0.20 2.1 3.63 ± 0.16 4.5 4.07 ± 0.15 3.6 4.31 ± 0.25 5.8 4.15 ± 0.23 5.5