Table II.
Limit loads of the five internal fixation techniques.
Internal fixation technique | n | Limit load, mean ± SD (N) |
---|---|---|
Circular wiring | 4 | 35.14±2.46 |
Figure of eight wiring | 4 | 56.61±1.97a |
Figure of eight plus circular wiring | 4 | 81.43±2.61b,c |
Kirschner tension band | 4 | 84.45±2.52b,c |
Screw fixation | 4 | 57.62±2.91a |
P<0.05 compared with circular wiring;
P<0.01 compared with circular wiring;
P<0.05 compared with figure of eight wiring or screw fixation.