Table I. Analytical performance levels (precision and accuracy) calculated for replicated calibration curves of ApoE4 peptide in ϵ4ϵ4 plasma diluted with ϵ3ϵ3 plasma.

Sample type n ϵ4ϵ4 plasma volume ratio Estimated concentration (μg/ml) CV (%)a Accuracyb
Standard 9 0.00 0.0 4.8 N/A
Standard 9 0.14 6.66 7.2 99.2
Standard 9 0.29 13.3 9.5 97.3
Standard 9 0.43 20.0 9.5 99.6
Standard 9 0.57 26.6 5.0 101.2
Standard 9 0.71 33.3 5.8 102.0
Standard 9 1.00 46.6 5.0 98.9
QC 9 0.17 7.99 6.6 99.7
QC 9 0.51 24.0 6.8 98.6
QC 9 0.86 40.0 7.2 102.9

a Expressed as the relative standard deviation (S.D.), i.e., S.D./mean × 100.

b Expressed as a percent difference: ((concentration found − concentration added)/concentration added) × 100.