Table 1.
Summary of Key Experimental Conditions and Results
PARAMETERS | Standard | Optimized | Oil-protected Dynamic Measurement |
---|---|---|---|
Time to Result | 2 hours | 30 min | 10 min |
Temperature | 37°C | Room Temp. | Room Temp. |
NaCl Concentration | 0.055M | 0.6M | 0.6M |
RESULTS | |||
Limit of Detection | 150 amole – 500 amole | 50 amole | 50 amoles |
Linear Dynamic Range | 150 amoles to 15 fmoles | 50 amoles to 15 fmoles | 50 amoles to 500 fmoles |