TABLE 1.

Accuracy and precision of stable hydrogen isotope ratio measurements by the platinum catalyzed H2-water equilibration method using international water reference materials1

References Accepted δ2H value Measured δ2H value2 Accuracy2 Mean accuracy Overall precision3
o/oo
GISP −189.7 (−190.2 to −189.1) −189.7 ± 1.3 −0.0 +0.7 1.5
−190.1 ± 1.7 −0.4
−188.8 ± 1.5 +0.9
IAEA 302A 506.2 (504.0–508.4) 504.8 ± 2.4 −1.4 −1.4 4.0
507.9 ± 3.4 +1.7
503.0 ± 4.4 −3.2
503.5 ± 5.4 −2.7
IAEA 302B 992.3 (985.6–999.0) 991.6 ± 2.4 −0.7 +2.8 2.7
995.4 ± 3.0 +3.1
998.7 ± 3.2 +6.4
994.8 ± 2.1 +2.5
1

Values are mean ± SD or accepted values (95%CI), n = 10. GISP, Greenland Ice Sheet Precipitation; 2H, deuterium; IAEA 302A, International Atomic Energy Agency 2H enriched reference water material 302A; IAEA 302B, International Atomic Energy Agency 2H enriched reference water material 302B.

2

Accuracy (o/oo) = measured δ2H value (o/oo) – accepted δ2H value (o/oo).

3

Overall precision (o/oo) = Inline graphic, where SD is the internal precision or SD of each set of measurements and n is the total number of sets of δ2H values used in the evaluation.