TABLE 5.
Troubleshooting table.
Step | Problem | Possible reason | Solution |
---|---|---|---|
8 | Δ R2 equal to 0 at all residue positions; DEST is unlikely to be observed if Δ R2 is not substantially greater than 0.2 s−1 | Insufficient population of dark state | Modify sample conditions or concentrations to increase the conversion rate from the NMR-visible to the dark species. This can often be accomplished by increasing the concentration of the dark state |
Negative values of Δ R2 are computed or R2 is not repeatable | Imperfect reference and dark-state sample condition matching. Small changes in pH (on the order of 0.2 pH units) near physiological pH of 7 can change the rate of 1HN exchange with 1H from bulk water on the order of 1 s−1 or greater | Option 1: the reference sample should be changed (i.e., pH adjusted) such that the sample conditions match. Resonance positions of 1H-15N correlations of histidine residues can be used as sensitive internal reporters of the relative pH of the two samples, and titrating the pH of the reference sample to precisely overlay with the 1H-15N correlation spectrum is a useful way to ensure sample consistency. If pH adjustment after sample creation is not possible, several reference samples with slightly different sample pH can be screened for the best overlay and/or used for R2 measurement. Option 2: pulse sequences that remove the contribution of water exchange to the measured relaxation rates12 can be used in place of standard 15N-R2 measurements; or lower the sample temperature |
|
10 | No DEST (off-resonance) saturation is observed | Saturation power is too weak | Increase the saturation power (Box 1) |
is too small to give rise to DEST | Increase by decreasing temperature or increasing the size of the dark state | ||
16 | Warning concerning local minima of optimization | It is difficult to guarantee a global optimum solution in an extremely large parameter space search | Convergence may be confirmed by independent fitting with different initial parameters, leading to the same solutions |
17 | Poor fit to DEST and/or ΔR2 data | Model is inappropriate for the system of interest | Change the fit type of the model (e.g., change from fit type 2, two-state model, to fit type 3 or 1, three-state or tethered/ bound models, respectively) |