. 2012 Nov;85(1019):1488–1498. doi: 10.1259/bjr/28015185

Table 1. Standard CT protocol parameters by age group for brain, chest and abdomen/pelvis examinations.

Examination Age group Set parameters
Voltage (kV) Qrefa (mAs) Rotation time (s) Pitch Detector configuration Beam collimation (mm)
CT brain 0–6 months 120 150 0.75 Axial 12×1.5 mm 18
6 months to 3 years 120 150 0.75 Axial 12×1.5 mm 18
3–6 years 120 200 0.75 Axial 12×1.5 mm 18
6–10 years 120 240 0.75 Axial 12×1.5 mm 18
>10 years 120 270 0.75 Axial 12×1.5 mm 18
CT chest <5 years 80 65 0.50 1.00 16×1.5 mm 24
5–10 years 100 80 0.50 1.00 16×1.5 mm 24
>10 years 120 80 0.50 1.00 16×1.5 mm 24
CT abdomen/pelvis <5 years 80 80 0.50 1.25 16×1.5 mm 24
5–10 years 100 80 0.50 1.25 16×1.5 mm 24
>10 years 120 60 0.50 1.25 16×1.5 mm 24

aQref is the imaging quality reference mAs, which is a setting specific to Siemens (Erlangen, Germany) that is used for automatic tube current modulation (CARE Dose4DTM; Siemens), which is defined by the user for each protocol. This value is adjusted based on image quality requirements and the amount of noise acceptable in the image. It is defined in terms of the effective mAs (actual mAs divided by pitch).