Table 1. Standard CT protocol parameters by age group for brain, chest and abdomen/pelvis examinations.
Examination | Age group | Set parameters |
|||||
Voltage (kV) | Qrefa (mAs) | Rotation time (s) | Pitch | Detector configuration | Beam collimation (mm) | ||
CT brain | 0–6 months | 120 | 150 | 0.75 | Axial | 12×1.5 mm | 18 |
6 months to 3 years | 120 | 150 | 0.75 | Axial | 12×1.5 mm | 18 | |
3–6 years | 120 | 200 | 0.75 | Axial | 12×1.5 mm | 18 | |
6–10 years | 120 | 240 | 0.75 | Axial | 12×1.5 mm | 18 | |
>10 years | 120 | 270 | 0.75 | Axial | 12×1.5 mm | 18 | |
CT chest | <5 years | 80 | 65 | 0.50 | 1.00 | 16×1.5 mm | 24 |
5–10 years | 100 | 80 | 0.50 | 1.00 | 16×1.5 mm | 24 | |
>10 years | 120 | 80 | 0.50 | 1.00 | 16×1.5 mm | 24 | |
CT abdomen/pelvis | <5 years | 80 | 80 | 0.50 | 1.25 | 16×1.5 mm | 24 |
5–10 years | 100 | 80 | 0.50 | 1.25 | 16×1.5 mm | 24 | |
>10 years | 120 | 60 | 0.50 | 1.25 | 16×1.5 mm | 24 |
aQref is the imaging quality reference mAs, which is a setting specific to Siemens (Erlangen, Germany) that is used for automatic tube current modulation (CARE Dose4DTM; Siemens), which is defined by the user for each protocol. This value is adjusted based on image quality requirements and the amount of noise acceptable in the image. It is defined in terms of the effective mAs (actual mAs divided by pitch).