Summary statistics for two manufactured panels showing the performance across and within platforms for their final verification values

Comparison and laboratory no.a Instrument no. Maximum CV Maximum SD Maximum bias SD bias % similarity CV
Reference panel comparison across platformsb
    Reference panel 1 22.5 0.1 0.32 0.1 2.9
    Reference panel 2 34.5 0.2 0.25 0.2 2.5
Panel 1 comparison within Roche instruments
    1 1 15.4 0.07 −0.21 0.11 1.1
2 23.2 0.1 −0.26 0.13 1.6
    2 3 26.1 0.11 0.09 0.08 1.1
4 30.2 0.14 −0.08 0.12 1.5
5 13.2 0.06 −0.14 0.07 0.9
6 30.1 0.15 −0.32 0.14 2.3
7 18 0.08 −0.12 0.13 1.5
8 17.8 0.08 −0.16 0.08 1.4
    3 9 36.3 0.16 −0.19 0.18 2.3
10 17.7 0.08 −0.16 0.09 1
    4 11 20.1 0.09 −0.14 0.1 1.7
12 25.9 0.1 −0.26 0.16 2
    5 13 25.9 0.1 −0.18 0.18 2
14 25.5 0.12 −0.11 0.12 1.4
    6 15 22.6 0.09 −0.17 0.12 1.4
16 22.2 0.1 0.12 0.1 1.6
    7 17 23.3 0.1 −0.15 0.1 1
    8 18 28.4 0.15 −0.13 0.13 0.9
19 19.4 0.09 −0.11 0.1 1.1
    Avg 23.2 0.1 −0.1 0.1 1.5
Panel 1 comparison within Abbott instruments
    2 20 18.1 0.08 −0.22 0.12 1.3
    9 21 13.3 0.06 0.15 0.08 1.2
22 24.1 0.1 −0.08 0.14 1.5
23 18.1 0.08 −0.1 0.07 1.1
24 36.4 0.17 0.09 0.15 1.6
    10 25 11.6 0.05 0.16 0.09 1.6
26 16.5 0.07 0.06 0.08 1
    11 27 23.8 0.12 0.12 0.12 1.6
28 19.1 0.08 0.1 0.09 1.2
    12 29 12.8 0.06 0.17 0.11 1.6
30 19 0.09 0.15 0.13 1.4
    13 31 26.6 0.1 −0.05 0.12 1.3
    14 32 20.5 0.09 0.06 0.11 1.2
    15 33 26.8 0.14 0.03 0.11 1.1
34 29.1 0.12 −0.06 0.17 1.5
    16 35 27.4 0.12 0.1 0.16 1.7
36 16.9 0.08 −0.11 0.09 1.1
    17 37 15.1 0.06 −0.1 0.08 0.9
38 33.2 0.13 0.11 0.15 1.7
    Avg 21.5 0.1 0.0 0.1 1.3
Panel 2 comparison within Roche instruments
    2 39 18.2 0.1 −0.25 0.19 2.4
    7 40 17 0.08 −0.19 0.13 1.8
41 15.7 0.1 −0.23 0.12 2
    Avg 17.0 0.1 −0.2 0.1 2.1
Panel 2 comparison within Abbott instruments
    17 42 23.4 0.11 0.12 0.14 1.7
43 16.1 0.07 0.11 0.14 1.6
    Avg 19.8 0.1 0.1 0.1 1.7
a

Numbers in the stub are laboratory numbers unless otherwise indicated (i.e., reference panel numbers).

b

Roche (n = 1) versus Abbott (n = 1).