Table 2.
Absolute average errors (±SD) for TR 40 ms and TR 100 ms as a function of flip angle FA. Absolute average errors vary systematically as a function of FA. For TR 40 ms, the optimal flip angle is 30°, with an absolute average percentage difference PD of 7.9% ± 3%. For a TR of 100 ms, the optimal flip angle is 60°, with an absolute average percentage difference PD of 10.2% ± 4.5%.
TR 40 ms | TR 100 ms | ||||
---|---|---|---|---|---|
FA | PD% | PD SD | FA | PD% | PD SD |
10° | 17.6 | 4.2 | 10° | 18.3 | 5.4 |
30° | 7.9 | 3 | 30° | 16.2 | 3.7 |
60° | 20.5 | 7.5 | 60° | 10.2 | 4.5 |
90° | 42 | 10.5 | 90° | 28.2 | 6.7 |